Electrical excitation beat-aided light-induced thermoelastic spectroscopy for high-robustness trace gas measurements

Heterodyne-based light-induced thermoelastic spectroscopy (HLITES) is capable of correcting measurement errors by evaluating the parameters of the quartz tuning fork (QTF). However, the correcting performance of HLITES will deteriorate under low concentration levels or laser power due to the weakened QTF transient response. Therefore, we propose an electrical excitation beat-aided LITES (EEBA-LITES), which is realized by optical and electrical excitation to the QTF utilizing the time-division multiplexing technique. Gas concentration and QTF parameters can be measured quasi-simultaneously. By normalizing the first harmonic (1<i>f</i>) signal with the background signal, a normalized 1<i>f</i>-signal was obtained to correct measurement errors resulting from power fluctuation and focus position change. The obtained nonlinear response of the normalized 1<i>f</i>-signal was used to correct the measurement error resulting from the resonant frequency shift. Unaffected by both gas concentration and laser power change, EEBA-LITES achieves the highest precision in beat-frequency-based HLITES and quartz-enhanced photoacoustic spectroscopy (QEPAS) techniques, with 1σ detection limits of ∼ 0.016 Hz for resonant frequency and ∼ 63 for quality factor for QTF, respectively. With the same hardware configuration, as the average laser power drops from 7.7 mW to 0.9 mW, EEBA-LITES enhances the detection limit of gas concentration and QTF resonant frequency by ∼ 3-7.8 times and ∼ 2.3-16.5 times, respectively, compared to previously reported self-correlated HLITES (SC-HLITES). Due to good resistance to interference factors, the EEBA-LITES demonstrates good potential for long-term, high-robustness field gas detection scenarios.

Authors

Publication Details

Journal
Optics Express
Published
2025-02-20
DOI
https://doi.org/10.1364/oe.557460
Citations
2
Primary Topic
Spectroscopy and Laser Applications
Type
article
Field-Weighted Citation Impact
1.51

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Electrical excitation beat-aided light-induced thermoelastic spectroscopy for high-robustness trace gas measurements

Xiaoteng Liu, Fang Song, Chuantao Zheng, Ziqi Zhao et al.
2 citations
Optics Express
Spectroscopy and Laser Applications
1.51
article

Electrical excitation beat-aided light-induced thermoelastic spectroscopy for high-robustness trace gas measurements

Xiaoteng Liu, Fang Song, Chuantao Zheng, Ziqi Zhao, Yu Zhang, Yiding Wang, Yanming Ma
article en
2 citations

Abstract

Heterodyne-based light-induced thermoelastic spectroscopy (HLITES) is capable of correcting measurement errors by evaluating the parameters of the quartz tuning fork (QTF). However, the correcting performance of HLITES will deteriorate under low concentration levels or laser power due to the weakened QTF transient response. Therefore, we propose an electrical excitation beat-aided LITES (EEBA-LITES), which is realized by optical and electrical excitation to the QTF utilizing the time-division multiplexing technique. Gas concentration and QTF parameters can be measured quasi-simultaneously. By normalizing the first harmonic (1<i>f</i>) signal with the background signal, a normalized 1<i>f</i>-signal was obtained to correct measurement errors resulting from power fluctuation and focus position change. The obtained nonlinear response of the normalized 1<i>f</i>-signal was used to correct the measurement error resulting from the resonant frequency shift. Unaffected by both gas concentration and laser power change, EEBA-LITES achieves the highest precision in beat-frequency-based HLITES and quartz-enhanced photoacoustic spectroscopy (QEPAS) techniques, with 1σ detection limits of ∼ 0.016 Hz for resonant frequency and ∼ 63 for quality factor for QTF, respectively. With the same hardware configuration, as the average laser power drops from 7.7 mW to 0.9 mW, EEBA-LITES enhances the detection limit of gas concentration and QTF resonant frequency by ∼ 3-7.8 times and ∼ 2.3-16.5 times, respectively, compared to previously reported self-correlated HLITES (SC-HLITES). Due to good resistance to interference factors, the EEBA-LITES demonstrates good potential for long-term, high-robustness field gas detection scenarios.

Optics ExpressVol. 33(5)
National Natural Science Foundation of China
Affordable and clean energy
Openalex Percentile: Top 21%
Spectroscopy and Laser Applications
1.51
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.