Laboratory evaluation of 4-channel irradiance radiometers suitable for observations on Argo floats

In this study, we evaluate the performance of three 4-channel irradiance radiometers suitable for the observations on Argo floats (Sea-Bird/Satlantic OCR-504, RBR quadrante, and SIAT SLIM-I4) through two laboratory experiments assessing their radiometric accuracy and stability, and their dark-signal stability and residual. Our findings highlight the importance of temperature correction for dark signals and provide insights into the aging drift effect over time. While all three sensors exhibit similar performance levels, the OCR-504 demonstrates high radiometric accuracy and stability but relatively low dark-signal stability; the quadrante offers excellent dark-signal stability, though with large dark-signal residual and relatively low radiometric accuracy and stability; and the SLIM-I4 excels in both radiometric and dark-signal performance. Additionally, we observed the aging drift on the dark signal of the OCR-504 over six years and found that the drift only affects the intercept term of dark signals, leaving the slope term unaffected. Based on these results, we recommend all the radiometers be equipped with an internal temperature sensor for more accurate temperature correction, and we propose three key technical criteria for irradiance radiometers on Argo floats: radiometric accuracy within ±3%, radiometric stability within 1% for irradiance and 0.1% for PAR, and dark-signal stability superior to 0.005 µW/cm<sup>2</sup>/nm for irradiance and 0.03 µmol/m<sup>2</sup>/s for PAR. These criteria are essential for ensuring reliable, consistent, and interoperable BGC-Argo datasets, as well as for facilitating the integration of new radiometers into the Argo program.

Authors

Institutions

Publication Details

Journal
Optics Express
Published
2025-02-20
DOI
https://doi.org/10.1364/oe.545359
Primary Topic
Oceanographic and Atmospheric Processes
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Laboratory evaluation of 4-channel irradiance radiometers suitable for observations on Argo floats

Shuguo Chen, Zhehai Shang, Xiaogang Xing, Ming Zhu
Optics Express
Oceanographic and Atmospheric Processes
article

Laboratory evaluation of 4-channel irradiance radiometers suitable for observations on Argo floats

Shuguo Chen, Zhehai Shang, Xiaogang Xing, Ming Zhu
article en

Abstract

In this study, we evaluate the performance of three 4-channel irradiance radiometers suitable for the observations on Argo floats (Sea-Bird/Satlantic OCR-504, RBR quadrante, and SIAT SLIM-I4) through two laboratory experiments assessing their radiometric accuracy and stability, and their dark-signal stability and residual. Our findings highlight the importance of temperature correction for dark signals and provide insights into the aging drift effect over time. While all three sensors exhibit similar performance levels, the OCR-504 demonstrates high radiometric accuracy and stability but relatively low dark-signal stability; the quadrante offers excellent dark-signal stability, though with large dark-signal residual and relatively low radiometric accuracy and stability; and the SLIM-I4 excels in both radiometric and dark-signal performance. Additionally, we observed the aging drift on the dark signal of the OCR-504 over six years and found that the drift only affects the intercept term of dark signals, leaving the slope term unaffected. Based on these results, we recommend all the radiometers be equipped with an internal temperature sensor for more accurate temperature correction, and we propose three key technical criteria for irradiance radiometers on Argo floats: radiometric accuracy within ±3%, radiometric stability within 1% for irradiance and 0.1% for PAR, and dark-signal stability superior to 0.005 µW/cm<sup>2</sup>/nm for irradiance and 0.03 µmol/m<sup>2</sup>/s for PAR. These criteria are essential for ensuring reliable, consistent, and interoperable BGC-Argo datasets, as well as for facilitating the integration of new radiometers into the Argo program.

Optics ExpressVol. 33(6)
Chinese Academy of Sciences (CN), Second Institute of Oceanography (CN), Shenzhen Institutes of Advanced Technology (CN), Ocean University of China (CN)
National Key Research and Development Program of China
Openalex Percentile: Top 98%
Oceanographic and Atmospheric Processes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.