Comparison of keratometry assessment between IOLMaster and Pentacam in dry eyes
Dry eye has significantly greater impact on keratometry measured with IOLMaster than with Pentacam. For cases with A-ACS between the two devices over 0.5 D, artificial tears are recommended before performing biometry.
Authors
- Jiayin Qin (ORCID: https://orcid.org/0000-0002-3777-5673)
- Fei Yang (ORCID: https://orcid.org/0000-0003-3213-1518)
- Xijuan Wang (ORCID: https://orcid.org/0000-0002-2409-115X)
- Mingwu Li (ORCID: https://orcid.org/0000-0002-1143-7027)
Institutions
- Peking University (CN)
- Peking University International Hospital (CN)
Publication Details
- Journal
- Photodiagnosis and Photodynamic Therapy
- Published
- 2025-02-20
- DOI
- https://doi.org/10.1016/j.pdpdt.2025.104527
- Citations
- 1
- Primary Topic
- Ocular Surface and Contact Lens
- Type
- article
- Field-Weighted Citation Impact
- 1.88