A Numerical Study on the Effect of Gate Position to the Structural Integrity of Plastic Injection-Molded Biomedical Implants

Optimizing the injection molding process for making biomedical implants is essential to avoid defects that impact patient safety and implant performance. This study examines how different gate positions affect defect rates in injection-molded polyether ether ketone (PEEK) pedicle screws, using numerical modeling to analyze melt flow, cavity pressure, sink mark depth, warpage, and Von Mises stress across four gate configurations. The findings show that gate position significantly influences structural integrity, with strategically placed gates reducing defects by promoting uniform flow. Gate configuration 2 was optimal, yielding minimal sink depth, low warpage, and low stress, while configuration 4 was the least effective as it was associated with higher values of defect rates. Carefully selecting gate positions can enhance the quality and reliability of biomedical implants.

Authors

Institutions

Publication Details

Published
2025-02-20
DOI
https://doi.org/10.3390/engproc2025084069
Primary Topic
Injection Molding Process and Properties
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Numerical Study on the Effect of Gate Position to the Structural Integrity of Plastic Injection-Molded Biomedical Implants

Edwell T. Mharakurwa, Abiodun Bayode, Fredrick Madaraka Mwema, Steven Otieno Otieno
2 links
Injection Molding Process and Properties
article

A Numerical Study on the Effect of Gate Position to the Structural Integrity of Plastic Injection-Molded Biomedical Implants

Edwell T. Mharakurwa, Abiodun Bayode, Fredrick Madaraka Mwema, Steven Otieno Otieno
article en

Abstract

Optimizing the injection molding process for making biomedical implants is essential to avoid defects that impact patient safety and implant performance. This study examines how different gate positions affect defect rates in injection-molded polyether ether ketone (PEEK) pedicle screws, using numerical modeling to analyze melt flow, cavity pressure, sink mark depth, warpage, and Von Mises stress across four gate configurations. The findings show that gate position significantly influences structural integrity, with strategically placed gates reducing defects by promoting uniform flow. Gate configuration 2 was optimal, yielding minimal sink depth, low warpage, and low stress, while configuration 4 was the least effective as it was associated with higher values of defect rates. Carefully selecting gate positions can enhance the quality and reliability of biomedical implants.

North-West University (ZA), Northumbria University (GB), Dedan Kimathi University of Technology (KE)
Openalex Percentile: Top 97%
Injection Molding Process and Properties
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A Numerical Study on the Effect of Gate Position to the Structural Integrity of Plastic Injection-Molded Biomedical Implants — Edwell T. Mharakurwa, Abiodun Bayode, et al. · (2025) | TGRS Research Map | TGRS