Two-layer calorimeter with thermally insulating separator for measuring ultrasonic power

Abstract The uncertainty in ultrasonic power measurement via calorimetry for high-power ultrasound is known to be larger than that via the radiation force balance (RFB) method. In order to achieve low uncertainty, we developed a two-layer calorimeter with a thermally insulating sheet to reduce the effect of heat generation by the ultrasonic transducer, which is a large factor in calorimetric uncertainty. The effect of the thickness of the separating sheet on ultrasound propagation and heat transfer was evaluated via simulations. A two-layer calorimeter with a 0.1-mm-thick thermally insulating sheet exhibited results close to those obtained via the RFB which is the most accurate method between 10 W–20 W. Also, we successfully measured ultrasonic power up to 100 W with this calorimeter. The results revealed that the two-layer calorimeter successfully suppressed the effect of heat generated by ultrasound transducers. This may lead to a reduction of the uncertainty in calorimetry.

Authors

Institutions

Publication Details

Journal
Japanese Journal of Applied Physics
Published
2025-02-20
DOI
https://doi.org/10.35848/1347-4065/adb855
Citations
1
Primary Topic
Scientific Measurement and Uncertainty Evaluation
Type
article
Field-Weighted Citation Impact
1.40
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Two-layer calorimeter with thermally insulating separator for measuring ultrasonic power

Takeyoshi Uchida, Choyu Uehara
1 citations
Japanese Journal of Applied Physics
Scientific Measurement and Uncertainty Evaluation
1.40
article

Two-layer calorimeter with thermally insulating separator for measuring ultrasonic power

Takeyoshi Uchida, Choyu Uehara
article en
1 citations

Abstract

Abstract The uncertainty in ultrasonic power measurement via calorimetry for high-power ultrasound is known to be larger than that via the radiation force balance (RFB) method. In order to achieve low uncertainty, we developed a two-layer calorimeter with a thermally insulating sheet to reduce the effect of heat generation by the ultrasonic transducer, which is a large factor in calorimetric uncertainty. The effect of the thickness of the separating sheet on ultrasound propagation and heat transfer was evaluated via simulations. A two-layer calorimeter with a 0.1-mm-thick thermally insulating sheet exhibited results close to those obtained via the RFB which is the most accurate method between 10 W–20 W. Also, we successfully measured ultrasonic power up to 100 W with this calorimeter. The results revealed that the two-layer calorimeter successfully suppressed the effect of heat generated by ultrasound transducers. This may lead to a reduction of the uncertainty in calorimetry.

Japanese Journal of Applied PhysicsVol. 64(3)
Shimane University Hospital (JP), National Institute of Advanced Industrial Science and Technology (JP)
Affordable and clean energy
Openalex Percentile: Top 19%
Scientific Measurement and Uncertainty Evaluation
1.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.