The Effect of Repeated Closing Torque on Torque Loss and Angular Deviation: An In Vitro Study

Loosening and fracturing of the central screw are common mechanical complications after implant restoration. However, the relationship between these complications and the repetitive tightening and loosening of the central screw during the fabrication and maintenance of the implant-supported restorations remains unknown. The purpose of this study was to evaluate the torque loss after repetitive tightening and loosening of the central screws on implants with different diameters as well as the changes in the angle deviation of the central screw relative to the implant. Twenty implants were divided into 2 groups based on diameter: 3.7 mm (group A) and 4.5 mm (group B) with 10 implants in each group. Each group was subdivided into 4 subgroups: A1, A2, B1, and B2 (n = 5). A closing torque of 15 N.cm was applied to groups A1 and B1, whereas a closing torque of 35 N.cm was applied to groups A2 and B2. Reverse torque measurements were taken 10 times for each group. The angular deviation of the central screw relative to the implant was recorded, and the surface wear of the central screw was observed under a scanning electron microscope. The data were analyzed using repeated measures 2-way analysis of variance (α = 0.05). Torque loss showed a significant upward trend across all groups with increased tightening cycle (P < .05). Implant diameter significantly influenced torque loss with smaller diameters exhibiting greater torque loss (P < .05). In addition, the angular deviation of the central screw relative to the implant was not affected by different diameters (P > .05). Still, it was affected by the closing torque and the cycles of multiple tightening and loosening procedures (P < .05). Under a 35 N.cm closing torque, initial torque loss ranged from 9.12 N.cm to 10.98 N.cm. Peak torque loss occurred at the 10th cycle with 16.40 N.cm values for 3.7-mm implants and 12.42 N.cm for 4.5-mm implants. Repeated tightening and loosening procedures increased both torque loss and angular deviation. The diameter of the implant may impact the torque loss with a larger diameter showing less torque loss. To reduce the risk of potential complications, it is suggested that the number of tightening cycles for narrow-diameter implants be limited.

Authors

Institutions

Publication Details

Journal
Journal of Oral Implantology
Published
2025-02-20
DOI
https://doi.org/10.1563/aaid-joi-d-24-00143
Citations
1
Primary Topic
Muscle activation and electromyography studies
Type
article
Field-Weighted Citation Impact
0.67
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Effect of Repeated Closing Torque on Torque Loss and Angular Deviation: An In Vitro Study

Ruoyan Guo, Guofu Lin, Run Chen, Yuehong Qian
1 citations
Journal of Oral Implantology
Muscle activation and electromyography studies
0.67
article

The Effect of Repeated Closing Torque on Torque Loss and Angular Deviation: An In Vitro Study

Ruoyan Guo, Guofu Lin, Run Chen, Yuehong Qian
article en
1 citations

Abstract

Loosening and fracturing of the central screw are common mechanical complications after implant restoration. However, the relationship between these complications and the repetitive tightening and loosening of the central screw during the fabrication and maintenance of the implant-supported restorations remains unknown. The purpose of this study was to evaluate the torque loss after repetitive tightening and loosening of the central screws on implants with different diameters as well as the changes in the angle deviation of the central screw relative to the implant. Twenty implants were divided into 2 groups based on diameter: 3.7 mm (group A) and 4.5 mm (group B) with 10 implants in each group. Each group was subdivided into 4 subgroups: A1, A2, B1, and B2 (n = 5). A closing torque of 15 N.cm was applied to groups A1 and B1, whereas a closing torque of 35 N.cm was applied to groups A2 and B2. Reverse torque measurements were taken 10 times for each group. The angular deviation of the central screw relative to the implant was recorded, and the surface wear of the central screw was observed under a scanning electron microscope. The data were analyzed using repeated measures 2-way analysis of variance (α = 0.05). Torque loss showed a significant upward trend across all groups with increased tightening cycle (P < .05). Implant diameter significantly influenced torque loss with smaller diameters exhibiting greater torque loss (P < .05). In addition, the angular deviation of the central screw relative to the implant was not affected by different diameters (P > .05). Still, it was affected by the closing torque and the cycles of multiple tightening and loosening procedures (P < .05). Under a 35 N.cm closing torque, initial torque loss ranged from 9.12 N.cm to 10.98 N.cm. Peak torque loss occurred at the 10th cycle with 16.40 N.cm values for 3.7-mm implants and 12.42 N.cm for 4.5-mm implants. Repeated tightening and loosening procedures increased both torque loss and angular deviation. The diameter of the implant may impact the torque loss with a larger diameter showing less torque loss. To reduce the risk of potential complications, it is suggested that the number of tightening cycles for narrow-diameter implants be limited.

Journal of Oral ImplantologyVol. 51(2)
Fujian Medical University (CN)
Openalex Percentile: Top 35%
Muscle activation and electromyography studies
0.67
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.