Calibration method for high-frequency electrotome quality analyzer based on a specific device

Abstract High frequency electrotome quality analyzer is widely used by medical institutions, medical equipment manufacturers, and medical equipment testing institutions. There are several important technical indicators, such as high-frequency power, high-frequency leakage current, and output resistance, to evaluate the metrological performance of high-frequency electrotome quality analyzers, but currently, there are no corresponding verification regulations or calibration specifications. This article presents a calibration method for high frequency electrotome quality analyzers by selecting a specific test device, and the rationality of the method is evaluated by verification experiments, which are conducted on typical types of quality analyzers manufactured by mainstream manufacturers. The experimental results are compared with the requirements stipulated by the national calibration specification JJF 1217-2009 Calibration Specification for Electrosurgical Generator to determine the feasibility of the presented calibration method.

Authors

Institutions

Publication Details

Journal
Journal of Physics Conference Series
Published
2025-02-01
DOI
https://doi.org/10.1088/1742-6596/2951/1/012076
Primary Topic
Fault Detection and Control Systems
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Calibration method for high-frequency electrotome quality analyzer based on a specific device

Chuandong Xie, Yanxiang Fu, Wei Liu, Tingting Ren et al.
Journal of Physics Conference Series
Fault Detection and Control Systems
article

Calibration method for high-frequency electrotome quality analyzer based on a specific device

Chuandong Xie, Yanxiang Fu, Wei Liu, Tingting Ren, Xu Yang
article en

Abstract

Abstract High frequency electrotome quality analyzer is widely used by medical institutions, medical equipment manufacturers, and medical equipment testing institutions. There are several important technical indicators, such as high-frequency power, high-frequency leakage current, and output resistance, to evaluate the metrological performance of high-frequency electrotome quality analyzers, but currently, there are no corresponding verification regulations or calibration specifications. This article presents a calibration method for high frequency electrotome quality analyzers by selecting a specific test device, and the rationality of the method is evaluated by verification experiments, which are conducted on typical types of quality analyzers manufactured by mainstream manufacturers. The experimental results are compared with the requirements stipulated by the national calibration specification JJF 1217-2009 Calibration Specification for Electrosurgical Generator to determine the feasibility of the presented calibration method.

Journal of Physics Conference SeriesVol. 2951(1)
Chongqing Metrology Quality Inspection and Research Institute (CN)
Openalex Percentile: Top 97%
Fault Detection and Control Systems
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Calibration method for high-frequency electrotome quality analyzer based on a specific device — Chuandong Xie, Yanxiang Fu, et al. · Journal of Physics Conference Series (2025) | TGRS Research Map | TGRS