Performance of Unmodified Mechanical Ventilators With 2% Hydrogen Gas Mixtures

<b>Background:</b> Molecular hydrogen (H<sub>2</sub>) is a breathable gas that has been shown to have anti-oxidative, anti-inflammatory, and anti-apoptotic properties that may positively impact ischemia-reperfusion injury. The provision of 2% H<sub>2</sub> through unmodified mechanical ventilators may facilitate the clinical translation of H<sub>2</sub> as a therapeutic in critical illness. The effect of 2% H<sub>2</sub> on ventilator performance is unknown. <b>Methods:</b> Unmodified Maquet Servo-i, Maquet Servo-u, Dräger Evita Infinity V500, and Dräger Evita Babylog VN500 ventilators from clinical stock were tested in an experimental closed system using certified, premixed air and O<sub>2</sub> containing 2% H<sub>2</sub> gas. Wall air and O<sub>2</sub> supply were used as control. Ventilator settings were varied across the spectrum of neonatal to adult settings. End points included (1) difference between set and delivered tidal volume (V<sub>T</sub>) (Douglas method), (2) difference between set versus delivered O<sub>2</sub> concentration, (3) delivered H<sub>2</sub> concentration (gas chromatography), and (4) ventilator pre-use check malfunction. Correlation between set and measured end points were quantified by linear regression analysis and bias by Bland-Altman analysis. <b>Results:</b> During H<sub>2</sub> administration, the average bias in measured versus set V<sub>T</sub> was within ± 10% for all ventilators except for the Babylog VN500, which exhibited an average bias of -89.2% (95% CI -107.0 to -71.3). The average bias in measured F<sub>IO<sub>2</sub></sub> was within ± 10% of set for all ventilators. Except for the Babylog VN500, all ventilators passed the pre-use check. <b>Conclusions:</b> Unmodified Servo-i, Servo-u, and Evita V500 ventilators deliver 2% H<sub>2</sub> mixtures with acceptable accuracy in V<sub>T</sub> and F<sub>IO<sub>2</sub></sub>. The Babylog VN500, which uses hot-wire anometry and a higher set operating temperature, exhibits unacceptably inaccurate delivery of V<sub>T</sub> with H<sub>2</sub> mixtures.

Authors

Institutions

Publication Details

Journal
Respiratory Care
Published
2025-04-01
DOI
https://doi.org/10.4187/respcare.12371
Citations
4
Primary Topic
Respiratory Support and Mechanisms
Type
article
Field-Weighted Citation Impact
6.07
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance of Unmodified Mechanical Ventilators With 2% Hydrogen Gas Mixtures

Marc Anders, Richard Nguyen, Kristen Sack, John N. Kheir et al.
4 citations
Respiratory Care
Respiratory Support and Mechanisms
6.07
article

Performance of Unmodified Mechanical Ventilators With 2% Hydrogen Gas Mixtures

Marc Anders, Richard Nguyen, Kristen Sack, John N. Kheir, Dantin Jeramy Roddy, Julia García Mancebo, Yifeng Peng, Salvador Carrasco Sosa
article en
4 citations

Abstract

<b>Background:</b> Molecular hydrogen (H<sub>2</sub>) is a breathable gas that has been shown to have anti-oxidative, anti-inflammatory, and anti-apoptotic properties that may positively impact ischemia-reperfusion injury. The provision of 2% H<sub>2</sub> through unmodified mechanical ventilators may facilitate the clinical translation of H<sub>2</sub> as a therapeutic in critical illness. The effect of 2% H<sub>2</sub> on ventilator performance is unknown. <b>Methods:</b> Unmodified Maquet Servo-i, Maquet Servo-u, Dräger Evita Infinity V500, and Dräger Evita Babylog VN500 ventilators from clinical stock were tested in an experimental closed system using certified, premixed air and O<sub>2</sub> containing 2% H<sub>2</sub> gas. Wall air and O<sub>2</sub> supply were used as control. Ventilator settings were varied across the spectrum of neonatal to adult settings. End points included (1) difference between set and delivered tidal volume (V<sub>T</sub>) (Douglas method), (2) difference between set versus delivered O<sub>2</sub> concentration, (3) delivered H<sub>2</sub> concentration (gas chromatography), and (4) ventilator pre-use check malfunction. Correlation between set and measured end points were quantified by linear regression analysis and bias by Bland-Altman analysis. <b>Results:</b> During H<sub>2</sub> administration, the average bias in measured versus set V<sub>T</sub> was within ± 10% for all ventilators except for the Babylog VN500, which exhibited an average bias of -89.2% (95% CI -107.0 to -71.3). The average bias in measured F<sub>IO<sub>2</sub></sub> was within ± 10% of set for all ventilators. Except for the Babylog VN500, all ventilators passed the pre-use check. <b>Conclusions:</b> Unmodified Servo-i, Servo-u, and Evita V500 ventilators deliver 2% H<sub>2</sub> mixtures with acceptable accuracy in V<sub>T</sub> and F<sub>IO<sub>2</sub></sub>. The Babylog VN500, which uses hot-wire anometry and a higher set operating temperature, exhibits unacceptably inaccurate delivery of V<sub>T</sub> with H<sub>2</sub> mixtures.

Respiratory CareVol. 70(4)
Boston Children's Hospital (US), Texas Children's Hospital (US)
Openalex Percentile: Top 4%
Respiratory Support and Mechanisms
6.07
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.