A Computational Study on the Neck‐Stem Rectangular Tapered Connection: Effects of Angular Mismatch, Assembly, and Cyclic Loading

The bi-modular hip prosthesis is characterized by two tapered connections: a circular cross-section at the head-neck interface and a rectangular cross-section at the neck-stem interface. Even if the latter guarantees customization, it concerns a high rate of early failure. The connection resistance is relatable to machining (tolerances cause angular mismatch), implantation (hammering force or manual), and usage (Body Mass Index [BMI]). Due to the lack of literature about the neck-stem coupling, this work aims to investigate how the geometry of the rectangular taper connection and the external loads affect the fatigue strength of a bi-modular hip prosthesis through a 3D Finite Element Model (FEM). Nine combinations of neck-stem coupling are obtained considering the tolerances' limits on frontal and lateral angles as <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"> <mml:semantics><mml:mrow><mml:mn>4</mml:mn> <mml:mo>°</mml:mo> <mml:msup><mml:mspace></mml:mspace> <mml:mrow> <mml:mtable> <mml:mtr><mml:mtd><mml:mo>+</mml:mo> <mml:msup><mml:mn>6</mml:mn> <mml:mo>'</mml:mo></mml:msup> </mml:mtd> </mml:mtr> <mml:mtr> <mml:mtd><mml:msup><mml:mn>0</mml:mn> <mml:mo>'</mml:mo></mml:msup> </mml:mtd> </mml:mtr> </mml:mtable> </mml:mrow> </mml:msup> </mml:mrow> <mml:annotation>$$ {4}^{{}^{\circ}\begin{array}{c}+{6}^{\prime}\\ {}{0}^{\prime}\end{array}} $$</mml:annotation></mml:semantics> </mml:math> . The CoCr neck and the Ti6Al4V stem, studied in their halved, are constrained and loaded inspired by the standard ISO 7206: the stem is distally encastered simulating the embedding and tilted by 10° concerning the sagittal plane, while the force is applied vertically. First, the influence of the assembly is investigated using <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"> <mml:semantics><mml:mrow><mml:mn>0.3</mml:mn> <mml:mspace></mml:mspace> <mml:mi>kN</mml:mi></mml:mrow> <mml:annotation>$$ 0.3\ \mathrm{kN} $$</mml:annotation></mml:semantics> </mml:math> , <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"> <mml:semantics><mml:mrow><mml:mn>2</mml:mn> <mml:mspace></mml:mspace> <mml:mi>kN</mml:mi></mml:mrow> <mml:annotation>$$ 2\ \mathrm{kN} $$</mml:annotation></mml:semantics> </mml:math> , and <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"> <mml:semantics><mml:mrow><mml:mn>4</mml:mn> <mml:mspace></mml:mspace> <mml:mi>kN</mml:mi></mml:mrow> <mml:annotation>$$ 4\ \mathrm{kN} $$</mml:annotation></mml:semantics> </mml:math> ; then, a cyclical vertical force varying from <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"> <mml:semantics><mml:mrow><mml:mn>2.67</mml:mn> <mml:mspace></mml:mspace> <mml:mi>kN</mml:mi></mml:mrow> <mml:annotation>$$ 2.67\ \mathrm{kN} $$</mml:annotation></mml:semantics> </mml:math> to <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"> <mml:semantics><mml:mrow><mml:mn>5.34</mml:mn> <mml:mspace></mml:mspace> <mml:mi>kN</mml:mi></mml:mrow> <mml:annotation>$$ 5.34\ \mathrm{kN} $$</mml:annotation></mml:semantics> </mml:math> is imposed. Finally, one combination is analyzed in its integrity to evaluate the effect of the out-of-plane load. The study's findings concern: (i) a positive angular mismatch, which is responsible for proximal contacts, improves fatigue life, reducing Sines stress up to 33%; (ii) the higher the assembly force the higher the neck stability and the lower the extension of the overstressed lateral area; (iii) the implant fatigue resistance is directly proportional to the patient's BMI; and (iv) the out-of-plane external load causes a 40% increment in the fatigue failure risk.

Authors

Institutions

Publication Details

Journal
International Journal for Numerical Methods in Biomedical Engineering
Published
2025-02-01
DOI
https://doi.org/10.1002/cnm.3909
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Computational Study on the Neck‐Stem Rectangular Tapered Connection: Effects of Angular Mismatch, Assembly, and Cyclic Loading

Luca Ciriello, Giancarlo Pennati, Francesca Berti, Tomaso Villa et al.
International Journal for Numerical Methods in Biomedical Engineering
Orthopaedic implants and arthroplasty
article

A Computational Study on the Neck‐Stem Rectangular Tapered Connection: Effects of Angular Mismatch, Assembly, and Cyclic Loading

Luca Ciriello, Giancarlo Pennati, Francesca Berti, Tomaso Villa, Luigi La Barbera, R. Cromi
article en

Abstract

The bi-modular hip prosthesis is characterized by two tapered connections: a circular cross-section at the head-neck interface and a rectangular cross-section at the neck-stem interface. Even if the latter guarantees customization, it concerns a high rate of early failure. The connection resistance is relatable to machining (tolerances cause angular mismatch), implantation (hammering force or manual), and usage (Body Mass Index [BMI]). Due to the lack of literature about the neck-stem coupling, this work aims to investigate how the geometry of the rectangular taper connection and the external loads affect the fatigue strength of a bi-modular hip prosthesis through a 3D Finite Element Model (FEM). Nine combinations of neck-stem coupling are obtained considering the tolerances' limits on frontal and lateral angles as <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"> <mml:semantics><mml:mrow><mml:mn>4</mml:mn> <mml:mo>°</mml:mo> <mml:msup><mml:mspace></mml:mspace> <mml:mrow> <mml:mtable> <mml:mtr><mml:mtd><mml:mo>+</mml:mo> <mml:msup><mml:mn>6</mml:mn> <mml:mo>'</mml:mo></mml:msup> </mml:mtd> </mml:mtr> <mml:mtr> <mml:mtd><mml:msup><mml:mn>0</mml:mn> <mml:mo>'</mml:mo></mml:msup> </mml:mtd> </mml:mtr> </mml:mtable> </mml:mrow> </mml:msup> </mml:mrow> <mml:annotation>$$ {4}^{{}^{\circ}\begin{array}{c}+{6}^{\prime}\\ {}{0}^{\prime}\end{array}} $$</mml:annotation></mml:semantics> </mml:math> . The CoCr neck and the Ti6Al4V stem, studied in their halved, are constrained and loaded inspired by the standard ISO 7206: the stem is distally encastered simulating the embedding and tilted by 10° concerning the sagittal plane, while the force is applied vertically. First, the influence of the assembly is investigated using <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"> <mml:semantics><mml:mrow><mml:mn>0.3</mml:mn> <mml:mspace></mml:mspace> <mml:mi>kN</mml:mi></mml:mrow> <mml:annotation>$$ 0.3\ \mathrm{kN} $$</mml:annotation></mml:semantics> </mml:math> , <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"> <mml:semantics><mml:mrow><mml:mn>2</mml:mn> <mml:mspace></mml:mspace> <mml:mi>kN</mml:mi></mml:mrow> <mml:annotation>$$ 2\ \mathrm{kN} $$</mml:annotation></mml:semantics> </mml:math> , and <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"> <mml:semantics><mml:mrow><mml:mn>4</mml:mn> <mml:mspace></mml:mspace> <mml:mi>kN</mml:mi></mml:mrow> <mml:annotation>$$ 4\ \mathrm{kN} $$</mml:annotation></mml:semantics> </mml:math> ; then, a cyclical vertical force varying from <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"> <mml:semantics><mml:mrow><mml:mn>2.67</mml:mn> <mml:mspace></mml:mspace> <mml:mi>kN</mml:mi></mml:mrow> <mml:annotation>$$ 2.67\ \mathrm{kN} $$</mml:annotation></mml:semantics> </mml:math> to <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"> <mml:semantics><mml:mrow><mml:mn>5.34</mml:mn> <mml:mspace></mml:mspace> <mml:mi>kN</mml:mi></mml:mrow> <mml:annotation>$$ 5.34\ \mathrm{kN} $$</mml:annotation></mml:semantics> </mml:math> is imposed. Finally, one combination is analyzed in its integrity to evaluate the effect of the out-of-plane load. The study's findings concern: (i) a positive angular mismatch, which is responsible for proximal contacts, improves fatigue life, reducing Sines stress up to 33%; (ii) the higher the assembly force the higher the neck stability and the lower the extension of the overstressed lateral area; (iii) the implant fatigue resistance is directly proportional to the patient's BMI; and (iv) the out-of-plane external load causes a 40% increment in the fatigue failure risk.

International Journal for Numerical Methods in Biomedical EngineeringVol. 41(2)
Politecnico di Milano (IT)
Openalex Percentile: Top 95%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.