Multicenter analytical performance evaluation of the ST Genesia thrombin generation system

Available data on the potential use of thrombin generation (TG) assays in clinical practice is promising but larger studies involving several centers are needed to confirm this added-value for clinical purposes. The objective of this evaluation was to assess the analytical performances of the ST Genesia using STG-ThromboScreen, STG-BleedScreen, and STG-DrugScreen reagents across three centers to support its use in multicenter studies. Repeatability and reproducibility have been evaluated using commercial plasmas and quality control (QC) samples. Accuracy was assessed by calculating QC biases from the manufacturer's assigned values. Frozen plasma samples from healthy donors and patients having thrombophilia, hemorrhagic disorders, or taking anticoagulants allowed the comparison of TG parameters between the analyzers from two French centers. Repeatability and reproducibility CVs were respectively below 5% and 10% whatever the reagent and the sample used. The later fell under 6% after normalization. Mean biases between observed and assigned QC values provided by the manufacturer were less than 5% for most TG parameters. We found a good agreement for all TG parameters between the two evaluating centers. Relative bias was below 5% for all combination of parameters and reagents except for Peak height (+14.2%), ETP (+7.3%), and ETP inhibition (-12.0%) using STG-ThromboScreen. We found satisfactory repeatability, reproducibility, accuracy, and inter-laboratory variability for TG parameters using the three available reagents across three centers, supporting the use of ST Genesia in multicenter clinical trials.

Authors

Institutions

Publication Details

Journal
Annales de biologie clinique
Published
2025-02-01
DOI
https://doi.org/10.1684/abc.2025.1942
Citations
1
Primary Topic
Blood Coagulation and Thrombosis Mechanisms
Type
article
Field-Weighted Citation Impact
1.77
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Multicenter analytical performance evaluation of the ST Genesia thrombin generation system

Thomas Sinegre, Joffrey Feriel, Laurie Talon, Claire Hemar et al.
1 citations
Annales de biologie clinique
Blood Coagulation and Thrombosis Mechanisms
1.77
article

Multicenter analytical performance evaluation of the ST Genesia thrombin generation system

Thomas Sinegre, Joffrey Feriel, Laurie Talon, Claire Hemar, Aurélien Lebreton, Mélanie Lecomte
article en
1 citations

Abstract

Available data on the potential use of thrombin generation (TG) assays in clinical practice is promising but larger studies involving several centers are needed to confirm this added-value for clinical purposes. The objective of this evaluation was to assess the analytical performances of the ST Genesia using STG-ThromboScreen, STG-BleedScreen, and STG-DrugScreen reagents across three centers to support its use in multicenter studies. Repeatability and reproducibility have been evaluated using commercial plasmas and quality control (QC) samples. Accuracy was assessed by calculating QC biases from the manufacturer's assigned values. Frozen plasma samples from healthy donors and patients having thrombophilia, hemorrhagic disorders, or taking anticoagulants allowed the comparison of TG parameters between the analyzers from two French centers. Repeatability and reproducibility CVs were respectively below 5% and 10% whatever the reagent and the sample used. The later fell under 6% after normalization. Mean biases between observed and assigned QC values provided by the manufacturer were less than 5% for most TG parameters. We found a good agreement for all TG parameters between the two evaluating centers. Relative bias was below 5% for all combination of parameters and reagents except for Peak height (+14.2%), ETP (+7.3%), and ETP inhibition (-12.0%) using STG-ThromboScreen. We found satisfactory repeatability, reproducibility, accuracy, and inter-laboratory variability for TG parameters using the three available reagents across three centers, supporting the use of ST Genesia in multicenter clinical trials.

Annales de biologie cliniqueVol. 83(1)
Stago (France) (FR), Centre Hospitalier Universitaire de Clermont-Ferrand (FR), Centre Hospitalier de Valenciennes (FR)
Affordable and clean energy
Openalex Percentile: Top 19%
Blood Coagulation and Thrombosis Mechanisms
1.77
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.