Relationship between difference in endotracheal tube cuff area and airway area with minimum cuff pressure for adequate airway sealing: a prospective observational study

It is essential for clinicians to select the appropriate endotracheal tube to ensure effective airway management. However, an unmatched endotracheal tube cuff area to the airway area can lead to air or secretion leakage, even at the recommended cuff pressure of 20-30 cmH<sub>2</sub>O. The present multicenter prospective observational study aimed to determine the relationship between the difference in cuff area and airway area with the minimum cuff pressure to avoid airway leakage. Adult patients who underwent mechanical ventilation were assigned into three groups, with a minimum cuff pressure of < 20, 20-30, and > 30 cmH<sub>2</sub>O, respectively, in order to have adequate airway sealing. The primary outcome was the difference between the endotracheal tube cuff area and airway area (cuff-airway area difference) that was calculated for the three groups. A total of 284 patients were included, with the mean age of 65.19 (± 14.03) years old. There were 166, 63 and 55 patients who required a minimum cuff pressure of < 20, 20-30 and > 30 cmH<sub>2</sub>O, respectively. The mean cuff-airway area difference was 236.00 ± 85.26, 149.70 ± 48.34 and - 12.29 ± 113.0 mm<sup>2</sup> in the < 20, 20-30, and > 30 cmH<sub>2</sub>O groups, respectively. In addition, the simple linear regression analysis revealed a negative linear relationship between the cuff-airway area difference and minimum cuff pressure (Y = -0.1266 × X + 46.50, F = 571.40, p < 0.001). It can be concluded that a significant number of patients require a cuff pressure out of the recommended range (< 20 or > 30 mmH<sub>2</sub>O) to have adequate airway sealing. Patients with a lower cuff-airway area difference require a higher minimum cuff pressure to seal the airway.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-02-18
DOI
https://doi.org/10.1038/s41598-025-85355-x
Citations
2
Primary Topic
Airway Management and Intubation Techniques
Type
article
Field-Weighted Citation Impact
4.38

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Relationship between difference in endotracheal tube cuff area and airway area with minimum cuff pressure for adequate airway sealing: a prospective observational study

Wang‐Qin Shen, Yue‐Hong Wu, Yanghui Xu, Jiahai Shi et al.
2 citations
Scientific Reports
Airway Management and Intubation Techniques
4.38
article

Relationship between difference in endotracheal tube cuff area and airway area with minimum cuff pressure for adequate airway sealing: a prospective observational study

Wang‐Qin Shen, Yue‐Hong Wu, Yanghui Xu, Jiahai Shi, Meijuan Lan, Hongwu Shen, Honglei Wu, Yan‐Ping Zhu, Lei Ding
article en
2 citations

Abstract

It is essential for clinicians to select the appropriate endotracheal tube to ensure effective airway management. However, an unmatched endotracheal tube cuff area to the airway area can lead to air or secretion leakage, even at the recommended cuff pressure of 20-30 cmH<sub>2</sub>O. The present multicenter prospective observational study aimed to determine the relationship between the difference in cuff area and airway area with the minimum cuff pressure to avoid airway leakage. Adult patients who underwent mechanical ventilation were assigned into three groups, with a minimum cuff pressure of < 20, 20-30, and > 30 cmH<sub>2</sub>O, respectively, in order to have adequate airway sealing. The primary outcome was the difference between the endotracheal tube cuff area and airway area (cuff-airway area difference) that was calculated for the three groups. A total of 284 patients were included, with the mean age of 65.19 (± 14.03) years old. There were 166, 63 and 55 patients who required a minimum cuff pressure of < 20, 20-30 and > 30 cmH<sub>2</sub>O, respectively. The mean cuff-airway area difference was 236.00 ± 85.26, 149.70 ± 48.34 and - 12.29 ± 113.0 mm<sup>2</sup> in the < 20, 20-30, and > 30 cmH<sub>2</sub>O groups, respectively. In addition, the simple linear regression analysis revealed a negative linear relationship between the cuff-airway area difference and minimum cuff pressure (Y = -0.1266 × X + 46.50, F = 571.40, p < 0.001). It can be concluded that a significant number of patients require a cuff pressure out of the recommended range (< 20 or > 30 mmH<sub>2</sub>O) to have adequate airway sealing. Patients with a lower cuff-airway area difference require a higher minimum cuff pressure to seal the airway.

Scientific ReportsVol. 15(1)
Nantong University (CN), Anhui University of Traditional Chinese Medicine (CN), Affiliated Hospital of Nantong University (CN), Zhongda Hospital Southeast University (CN), Second Affiliated Hospital of Zhejiang University (CN)
Government of Jiangsu Province, Nantong University
Openalex Percentile: Top 8%
Airway Management and Intubation Techniques
4.38
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.