Force Sensor for Instrumented Patellar Prostheses: Development and Characterization

The development of an instrumented patellar prosthesis, able to measure the contact forces at the patellofemoral joint, can significantly aid in investigating the causes of total knee arthroplasty failures due to patellar complications. This study focuses on developing and validating an instrumented patellar prosthesis to measure contact forces in the patellofemoral joint. A piezoresistive force sensor was characterized and integrated into a conditioning circuit, with the aim of its implementation in the prosthesis. To measure medial and lateral forces independently, the sensors were trimmed in half. Compression tests (up to 2000 N) assessed sensor performance in terms of linearity (R<sup>2</sup> = 0.998 intact vs. 0.989 trimmed), repeatability (0.9% intact vs. 0.8% trimmed), and accuracy (1.7% intact vs. 2.3% trimmed) for forces up to 250 N. Higher force levels resulted in increased errors, but at a rate still comparable to that of existing sensors in the literature. Key considerations for the design of the instrumented prosthesis, such as minimizing point and shear loads, were identified. A prototype prosthesis capable of housing the sensor was proposed. The integrated system shows potential for improving the understanding of Total knee arthroplasty (TKA) failures through in vitro studies and could serve as an intraoperative tool for the evaluation of bone resections.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2025-02-18
DOI
https://doi.org/10.3390/s25041226
Citations
4
Primary Topic
Total Knee Arthroplasty Outcomes
Type
article
Field-Weighted Citation Impact
8.03
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Force Sensor for Instrumented Patellar Prostheses: Development and Characterization

Vera Maioli, Matteo Zauli, Angelo Cappello, Luca Cristofolini
4 citations
Sensors
Total Knee Arthroplasty Outcomes
8.03
article

Force Sensor for Instrumented Patellar Prostheses: Development and Characterization

Vera Maioli, Matteo Zauli, Angelo Cappello, Luca Cristofolini
article en
4 citations

Abstract

The development of an instrumented patellar prosthesis, able to measure the contact forces at the patellofemoral joint, can significantly aid in investigating the causes of total knee arthroplasty failures due to patellar complications. This study focuses on developing and validating an instrumented patellar prosthesis to measure contact forces in the patellofemoral joint. A piezoresistive force sensor was characterized and integrated into a conditioning circuit, with the aim of its implementation in the prosthesis. To measure medial and lateral forces independently, the sensors were trimmed in half. Compression tests (up to 2000 N) assessed sensor performance in terms of linearity (R<sup>2</sup> = 0.998 intact vs. 0.989 trimmed), repeatability (0.9% intact vs. 0.8% trimmed), and accuracy (1.7% intact vs. 2.3% trimmed) for forces up to 250 N. Higher force levels resulted in increased errors, but at a rate still comparable to that of existing sensors in the literature. Key considerations for the design of the instrumented prosthesis, such as minimizing point and shear loads, were identified. A prototype prosthesis capable of housing the sensor was proposed. The integrated system shows potential for improving the understanding of Total knee arthroplasty (TKA) failures through in vitro studies and could serve as an intraoperative tool for the evaluation of bone resections.

SensorsVol. 25(4)
University of Bologna (IT)
Openalex Percentile: Top 3%
Total Knee Arthroplasty Outcomes
8.03
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Force Sensor for Instrumented Patellar Prostheses: Development and Characterization — Vera Maioli, Matteo Zauli, et al. · Sensors (2025) | TGRS Research Map | TGRS