Computed Tomography Doses Calculation: Do We Really Need a New Dose Assessment Tool?

<b>Background/Objectives:</b> The increasing use of computed tomography (CT) scans significantly contributes to population exposure to ionizing radiation. Traditional dose metrics, such as dose-length product (DLP) and effective dose (ED), lack precision in reflecting individual radiation exposure. This study introduces a novel parameters such as size-specific effective dose (EDss) and the size-specific dose-length product (DLPss), to improve patient-specific dose estimation. The aim of this study is to enhance dose calculation accuracy, optimize CT protocols, and guide the development of next-generation CT technologies. <b>Methods:</b> A retrospective analysis of 247 abdominal and pelvic CT scans (113 women, 134 men) was conducted. Anthropometric parameters, including body mass index (BMI), cross-sectional dimensions, and dose indices, were measured. EDss and DLPss were calculated using size-specific correction factors, and statistical correlations between these parameters were assessed. <b>Results:</b> The mean BMI was 25.92 ± 5.34. DLPss values ranged from 261.63 to 1217.70 mGy·cm (mean: 627.83 ± 145.32) and were roughly 21% higher than traditional DLP values, with men showing slightly higher mean values than women. EDss values ranged from 6.65 to 15.45 mSv (mean: 9.42 ± 2.18 mSv), approximately 22% higher than traditional ED values, demonstrating improved individualization. Significant correlations were observed between BMI and effective diameter (r = 0.78), with stronger correlations in men (r = 0.85). The mean CTDIvol was 11.37 ± 3.50 mGy, and SSDE averaged 13.91 ± 2.39 mGy. Scan length reductions were observed in 53.8% of cases, with statistically significant differences by gender. <b>Conclusions:</b> EDss and DLPss offer improved accuracy in radiation dose estimation, addressing the limitations of traditional methods. Their adoption into clinical protocols, supported by AI-driven automation, could optimize diagnostic safety and significantly reduce radiation risk for patients. Further multicenter studies and technological advancements are recommended to validate these metrics and facilitate their integration into daily practice.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2025-02-18
DOI
https://doi.org/10.3390/jcm14041348
Citations
1
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
1.36
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Computed Tomography Doses Calculation: Do We Really Need a New Dose Assessment Tool?

Maciej Piskunowicz, Magdalena Wszędybył-Winklewska, Małgorzata Grzywińska, Arkadiusz Szarmach et al.
1 citations
Journal of Clinical Medicine
Radiation Dose and Imaging
1.36
article

Computed Tomography Doses Calculation: Do We Really Need a New Dose Assessment Tool?

Maciej Piskunowicz, Magdalena Wszędybył-Winklewska, Małgorzata Grzywińska, Arkadiusz Szarmach, Paweł Gać, Dominika Sabiniewicz-Ziajka
article en
1 citations

Abstract

<b>Background/Objectives:</b> The increasing use of computed tomography (CT) scans significantly contributes to population exposure to ionizing radiation. Traditional dose metrics, such as dose-length product (DLP) and effective dose (ED), lack precision in reflecting individual radiation exposure. This study introduces a novel parameters such as size-specific effective dose (EDss) and the size-specific dose-length product (DLPss), to improve patient-specific dose estimation. The aim of this study is to enhance dose calculation accuracy, optimize CT protocols, and guide the development of next-generation CT technologies. <b>Methods:</b> A retrospective analysis of 247 abdominal and pelvic CT scans (113 women, 134 men) was conducted. Anthropometric parameters, including body mass index (BMI), cross-sectional dimensions, and dose indices, were measured. EDss and DLPss were calculated using size-specific correction factors, and statistical correlations between these parameters were assessed. <b>Results:</b> The mean BMI was 25.92 ± 5.34. DLPss values ranged from 261.63 to 1217.70 mGy·cm (mean: 627.83 ± 145.32) and were roughly 21% higher than traditional DLP values, with men showing slightly higher mean values than women. EDss values ranged from 6.65 to 15.45 mSv (mean: 9.42 ± 2.18 mSv), approximately 22% higher than traditional ED values, demonstrating improved individualization. Significant correlations were observed between BMI and effective diameter (r = 0.78), with stronger correlations in men (r = 0.85). The mean CTDIvol was 11.37 ± 3.50 mGy, and SSDE averaged 13.91 ± 2.39 mGy. Scan length reductions were observed in 53.8% of cases, with statistically significant differences by gender. <b>Conclusions:</b> EDss and DLPss offer improved accuracy in radiation dose estimation, addressing the limitations of traditional methods. Their adoption into clinical protocols, supported by AI-driven automation, could optimize diagnostic safety and significantly reduce radiation risk for patients. Further multicenter studies and technological advancements are recommended to validate these metrics and facilitate their integration into daily practice.

Journal of Clinical MedicineVol. 14(4)
Wroclaw Medical University (PL), Uniwersytet Pomorski w Słupsku (PL), Gdańsk Medical University (PL)
Openalex Percentile: Top 23%
Radiation Dose and Imaging
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.