Design and Simulation of a Bioimpedance Sensor for Measuring the Electrical Properties of Blood in the Hemodialysis Process

This study presents the design and simulation of a bioimpedance sensor aimed at measuring the electrical properties of blood during the hemodialysis process. Utilizing two configurations—4-electrode washer and Van der Pauw—the simulations were conducted in the COMSOL environment using the finite element method. Performance accuracy was evaluated against data from Gabriel et al., with results analyzed in MATLAB, including Nyquist curve examination under various conditions. The findings indicate that for the 4-electrode washer method, optimal sensor performance occurs when the current injection radius matches the voltage reading radius; however, this configuration presents construction challenges. Conversely, the 4-electrode Van der Pauw method demonstrated that the radius of the electrodes significantly impacts sensor evaluation and performance. By optimizing electrode dimensions, a compact sensor design can be developed, potentially enhancing the quality of dialysis machines and improving patient outcomes.

Authors

Institutions

Publication Details

Published
2025-02-04
DOI
https://doi.org/10.1109/aree63378.2025.10880299
Citations
2
Primary Topic
Body Composition Measurement Techniques
Type
article
Field-Weighted Citation Impact
2.33
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design and Simulation of a Bioimpedance Sensor for Measuring the Electrical Properties of Blood in the Hemodialysis Process

Rasool Baghbani, Masoomeh Ashoorirad, Adel Reyhanian
2 citations
Body Composition Measurement Techniques
2.33
article

Design and Simulation of a Bioimpedance Sensor for Measuring the Electrical Properties of Blood in the Hemodialysis Process

Rasool Baghbani, Masoomeh Ashoorirad, Adel Reyhanian
article en
2 citations

Abstract

This study presents the design and simulation of a bioimpedance sensor aimed at measuring the electrical properties of blood during the hemodialysis process. Utilizing two configurations—4-electrode washer and Van der Pauw—the simulations were conducted in the COMSOL environment using the finite element method. Performance accuracy was evaluated against data from Gabriel et al., with results analyzed in MATLAB, including Nyquist curve examination under various conditions. The findings indicate that for the 4-electrode washer method, optimal sensor performance occurs when the current injection radius matches the voltage reading radius; however, this configuration presents construction challenges. Conversely, the 4-electrode Van der Pauw method demonstrated that the radius of the electrodes significantly impacts sensor evaluation and performance. By optimizing electrode dimensions, a compact sensor design can be developed, potentially enhancing the quality of dialysis machines and improving patient outcomes.

Hamedan University of Technology (IR)
Affordable and clean energy
Openalex Percentile: Top 14%
Body Composition Measurement Techniques
2.33
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.