Dose-Optimized Image Acquisition Parameters for Neonatal Chest Radiography: A Phantom Study Comparing Computed Radiography and Wireless Digital Radiography Needle Detectors

To determine dose-optimized image acquisition parameters for good image quality (IQ) in neonatal chest radiography with a computed radiography (CR) CsBr needle detector vs. a wireless digital radiography (DR) CsI detector using different doses and filters.Physical resolution of the two detectors in unprocessed imaging of a contrast-detail phantom was automatically evaluated. Post-processed chest radiographic imaging of a neonatal phantom was used for Visual Grading Analysis (VGA) by three radiology raters. Different kVp, mAs, and filter settings were used. The VGA score (VGAS) and dose area product (DAP) were used to determine image acquisition parameters and dose levels for good image quality. Pixel data from segments corresponding to visual grading characteristics (VGC) was used to calculate signal-to-noise ratio, contrast-to-noise ratio (CNR), and signal profile curves. These results were compared to the raters' "ground truth" by Spearman's correlation.The CR detector had the highest resolution in unprocessed imaging, although this was dependent on a tube voltage of 66 kVp (P < 0.001), and more so than the DR detector. The VGAS showed no significant difference between the CR needle and the DR CsI detectors at the same DAP, or when using standard pediatric filtering of 3.5 mm Al + 0.1 mm Cu (P > 0.05). A lung dose level of 0.017 mSv was needed for good IQ (effective dose (<i>E</i>): 0.010 mSv). This was achievable with different acquisition parameters. Out of 24 segments, only the CNR of bone-to-soft-tissue had a good Spearman's correlation (ρ > 0.50) to raters' VGAS (P < 0.0001), mostly due to problems with image registration.The CR needle and DR CsI detectors have comparable IQ in neonatal chest radiography. In this study, an <i>E</i> of approximately 0.010 mSv was needed for good IQ. · CR needle and DR detectors have comparable image quality in neonatal chest radiography.. · CR needle technology has higher absolute raw image resolution, although this is voltage-dependent and more so than the DR detector.. · We propose ideal image acquisition parameters for neonatal chest radiography.. · Rama K, Esser M, Spogis J et al. Dose-Optimized Image Acquisition Parameters for Neonatal Chest Radiography: A Phantom Study Comparing Computed Radiography and Wireless Digital Radiography Needle Detectors. Rofo 2025; 197: 1416-1425.

Authors

Institutions

Publication Details

Journal
RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren
Published
2025-02-18
DOI
https://doi.org/10.1055/a-2525-9430
Citations
1
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
1.36

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Dose-Optimized Image Acquisition Parameters for Neonatal Chest Radiography: A Phantom Study Comparing Computed Radiography and Wireless Digital Radiography Needle Detectors

Jakob Spogis, Friedrich Wanninger, Bernd Hoberg, Michael Esser et al.
1 citations
RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren
Radiation Dose and Imaging
1.36
article

Dose-Optimized Image Acquisition Parameters for Neonatal Chest Radiography: A Phantom Study Comparing Computed Radiography and Wireless Digital Radiography Needle Detectors

Jakob Spogis, Friedrich Wanninger, Bernd Hoberg, Michael Esser, Jürgen Schäfer, K. Rama
article en
1 citations

Abstract

To determine dose-optimized image acquisition parameters for good image quality (IQ) in neonatal chest radiography with a computed radiography (CR) CsBr needle detector vs. a wireless digital radiography (DR) CsI detector using different doses and filters.Physical resolution of the two detectors in unprocessed imaging of a contrast-detail phantom was automatically evaluated. Post-processed chest radiographic imaging of a neonatal phantom was used for Visual Grading Analysis (VGA) by three radiology raters. Different kVp, mAs, and filter settings were used. The VGA score (VGAS) and dose area product (DAP) were used to determine image acquisition parameters and dose levels for good image quality. Pixel data from segments corresponding to visual grading characteristics (VGC) was used to calculate signal-to-noise ratio, contrast-to-noise ratio (CNR), and signal profile curves. These results were compared to the raters' "ground truth" by Spearman's correlation.The CR detector had the highest resolution in unprocessed imaging, although this was dependent on a tube voltage of 66 kVp (P < 0.001), and more so than the DR detector. The VGAS showed no significant difference between the CR needle and the DR CsI detectors at the same DAP, or when using standard pediatric filtering of 3.5 mm Al + 0.1 mm Cu (P > 0.05). A lung dose level of 0.017 mSv was needed for good IQ (effective dose (<i>E</i>): 0.010 mSv). This was achievable with different acquisition parameters. Out of 24 segments, only the CNR of bone-to-soft-tissue had a good Spearman's correlation (ρ > 0.50) to raters' VGAS (P < 0.0001), mostly due to problems with image registration.The CR needle and DR CsI detectors have comparable IQ in neonatal chest radiography. In this study, an <i>E</i> of approximately 0.010 mSv was needed for good IQ. · CR needle and DR detectors have comparable image quality in neonatal chest radiography.. · CR needle technology has higher absolute raw image resolution, although this is voltage-dependent and more so than the DR detector.. · We propose ideal image acquisition parameters for neonatal chest radiography.. · Rama K, Esser M, Spogis J et al. Dose-Optimized Image Acquisition Parameters for Neonatal Chest Radiography: A Phantom Study Comparing Computed Radiography and Wireless Digital Radiography Needle Detectors. Rofo 2025; 197: 1416-1425.

RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden VerfahrenVol. 197(12)
Agfa-Gevaert (Germany) (DE), University Children's Hospital Tübingen (DE)
European Maritime and Fisheries Fund
Openalex Percentile: Top 23%
Radiation Dose and Imaging
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.