High Risk of Revision Associated with the L-Cup Titanium Alloy Porous Coated Acetabular Component in Primary Total Hip Arthroplasty: Minimum Follow-Up of 14 Years

<b>Background:</b> Hip joint pain due to arthritis is a prevalent issue in adults, often necessitating surgical intervention such as total hip arthroplasty (THA). This procedure has been celebrated for its reliability; however, successful outcomes depend on numerous factors. Current advancements are focused on improving implant design and surgical methodologies. This study aimed to evaluate the long-term clinical and functional outcomes of uncemented total hip arthroplasty utilizing the L-Cup acetabular component. <b>Methods:</b> Between February 1999 and November 2010, 351 L-Cup components were implanted in 315 patients. A follow-up period ranged from 14 to 25 years. The clinical outcomes were assessed using the modified Merle d'Aubigné and Postel (MAP) classification and patient satisfaction was measured using a Visual Analog Scale (VAS). <b>Results:</b> Postoperative evaluations showed significant improvement, with VAS scores decreasing from a mean of 7.2 to 2.1, indicating substantial pain alleviation. The modified MAP classification showed a significant improvement of 6.3 points throughout the follow-up period. The results revealed that 49.5% of the cases were classified as excellent, while 20.5% had poor outcomes due to prosthesis loosening. According to the Kaplan-Meier estimator, the 5-year survival rate for the acetabular component was 97.78%, with survival rates of 90.5% at 10 years, 80.45% at 15 years, and 73.79% at 20 years. <b>Conclusions:</b> Total hip arthroplasty is an effective treatment for advanced degenerative joint diseases. While significant postoperative improvements were documented, the observed prosthesis loosening in 20.5% of cases raises concerns about the long-term effectiveness of the L-Cup acetabular component and suggests the need for further refinement in surgical techniques and implant design.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2025-02-15
DOI
https://doi.org/10.3390/jcm14041301
Citations
2
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
4.01
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

High Risk of Revision Associated with the L-Cup Titanium Alloy Porous Coated Acetabular Component in Primary Total Hip Arthroplasty: Minimum Follow-Up of 14 Years

Bartosz Gonera, Andrzej Borowski, Łukasz Olewnik, Marek Drobniewski et al.
2 citations
Journal of Clinical Medicine
Orthopaedic implants and arthroplasty
4.01
article

High Risk of Revision Associated with the L-Cup Titanium Alloy Porous Coated Acetabular Component in Primary Total Hip Arthroplasty: Minimum Follow-Up of 14 Years

Bartosz Gonera, Andrzej Borowski, Łukasz Olewnik, Marek Drobniewski, George Triantafyllou, Kacper Ruzik, Adam Borowski
article en
2 citations

Abstract

<b>Background:</b> Hip joint pain due to arthritis is a prevalent issue in adults, often necessitating surgical intervention such as total hip arthroplasty (THA). This procedure has been celebrated for its reliability; however, successful outcomes depend on numerous factors. Current advancements are focused on improving implant design and surgical methodologies. This study aimed to evaluate the long-term clinical and functional outcomes of uncemented total hip arthroplasty utilizing the L-Cup acetabular component. <b>Methods:</b> Between February 1999 and November 2010, 351 L-Cup components were implanted in 315 patients. A follow-up period ranged from 14 to 25 years. The clinical outcomes were assessed using the modified Merle d'Aubigné and Postel (MAP) classification and patient satisfaction was measured using a Visual Analog Scale (VAS). <b>Results:</b> Postoperative evaluations showed significant improvement, with VAS scores decreasing from a mean of 7.2 to 2.1, indicating substantial pain alleviation. The modified MAP classification showed a significant improvement of 6.3 points throughout the follow-up period. The results revealed that 49.5% of the cases were classified as excellent, while 20.5% had poor outcomes due to prosthesis loosening. According to the Kaplan-Meier estimator, the 5-year survival rate for the acetabular component was 97.78%, with survival rates of 90.5% at 10 years, 80.45% at 15 years, and 73.79% at 20 years. <b>Conclusions:</b> Total hip arthroplasty is an effective treatment for advanced degenerative joint diseases. While significant postoperative improvements were documented, the observed prosthesis loosening in 20.5% of cases raises concerns about the long-term effectiveness of the L-Cup acetabular component and suggests the need for further refinement in surgical techniques and implant design.

Journal of Clinical MedicineVol. 14(4)
National and Kapodistrian University of Athens (GR), Medical University of Warsaw (PL), Mazovian University in Płock (PL), Medical University of Lodz (PL)
Openalex Percentile: Top 9%
Orthopaedic implants and arthroplasty
4.01
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.