A Comparative Study of Pure and Silver-Enhanced Graphene Antennas for Fat-Intrabody Communication
This paper presents the design and comparative study of two implant antennas utilizing graphene material for fat-intrabody communication (fat-IBC) in the 5.8 GHz Industrial, Scientific, and Medical (ISM) band. Both antennas use pure graphene and silver-enhanced graphene as radiating elements, fabricated on a Taconic TLY-3 substrate. Copper-conducting elements were used as the substitute conductive material in CST Microwave Studio simulation for the optimized designs. These antennas were then measured in a fat-tissue-equivalent phantom model, and the S<inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">11</inf>reflection coefficients were recorded. These results show that the silver-enhanced graphene antenna has better performance in conductivity and resonance sharpness compared with the pure graphene antenna, which can be applied for the fat-IBC system.
Authors
- Mohamad Zoinol Abidin Abd Aziz (ORCID: https://orcid.org/0000-0002-9926-158X)
- Noor Badariah Asan (ORCID: https://orcid.org/0000-0002-6899-1424)
- Sree Aran Subramaniam
Institutions
- Technical University of Malaysia Malacca (MY)
Publication Details
- Published
- 2024-12-21
- DOI
- https://doi.org/10.1109/apace62360.2024.10877323
- Primary Topic
- Wireless Body Area Networks
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Ministry of Higher Education
- Universiti Teknikal Malaysia Melaka