Commissioning of a motion management system for a 1.5T Elekta Unity MR‐Linac: A single institution experience
MRI-guided target tracking and automated beam delivery control were successfully commissioned for the Elekta Unity MR-Linac. These gating features were shown to be highly accurate with an effectively small beam latency for a range of regular and irregular respiratory breathing traces.
Authors
- Daniel Ellis Hyer (ORCID: https://orcid.org/0000-0002-7638-9599)
- Joël St‐Aubin (ORCID: https://orcid.org/0000-0001-9356-8567)
- Blake R. Smith (ORCID: https://orcid.org/0009-0000-0117-2590)
Institutions
- University of Iowa (US)
Publication Details
- Journal
- Journal of Applied Clinical Medical Physics
- Published
- 2025-02-16
- DOI
- https://doi.org/10.1002/acm2.70005
- Citations
- 10
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 18.83