Verification of analytical performance of Microalbuminuria assay on the Abbott Alinity ci®: Experience of the central laboratory of Mohammed VI University Hospital of Oujda

The verification of analytical methods is crucial to ensure the accuracy of results. This study evaluates the performance of the microalbuminuria assay method in the central laboratory of CHU Mohammed VI in Oujda, to confirm its reliability for the diagnosis and monitoring of renal disorders and related conditions. We conducted a performance study of the Alinity ci® system by assessing repeatability and reproducibility in accordance with the COFRAC GTA 04 technical accreditation guidelines, in line with the quality requirements of the ISO 15189 standard. The results of our study indicate satisfactory repeatability for the two levels (low and high), with coefficients of variation of 1.24% and 1.12%, respectively. Additionally, reproducibility was also satisfactory, with coefficients of variation for these same levels of 4.73% and 4.75%, respectively. The reliability of the microalbuminuria assay results in our laboratory is confirmed by the satisfactory results obtained in our study, which comply with the recommendations of RICOS and FSCB.

Authors

Publication Details

Journal
GSC Biological and Pharmaceutical Sciences
Published
2025-02-16
DOI
https://doi.org/10.30574/gscbps.2025.30.2.0054
Primary Topic
Chronic Kidney Disease and Diabetes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Verification of analytical performance of Microalbuminuria assay on the Abbott Alinity ci®: Experience of the central laboratory of Mohammed VI University Hospital of Oujda

Mohammed Choukri, El Houcine Sebbar, Dounia Elmoujtahide, Yassir El Azraky
GSC Biological and Pharmaceutical Sciences
Chronic Kidney Disease and Diabetes
article

Verification of analytical performance of Microalbuminuria assay on the Abbott Alinity ci®: Experience of the central laboratory of Mohammed VI University Hospital of Oujda

Mohammed Choukri, El Houcine Sebbar, Dounia Elmoujtahide, Yassir El Azraky
article en

Abstract

The verification of analytical methods is crucial to ensure the accuracy of results. This study evaluates the performance of the microalbuminuria assay method in the central laboratory of CHU Mohammed VI in Oujda, to confirm its reliability for the diagnosis and monitoring of renal disorders and related conditions. We conducted a performance study of the Alinity ci® system by assessing repeatability and reproducibility in accordance with the COFRAC GTA 04 technical accreditation guidelines, in line with the quality requirements of the ISO 15189 standard. The results of our study indicate satisfactory repeatability for the two levels (low and high), with coefficients of variation of 1.24% and 1.12%, respectively. Additionally, reproducibility was also satisfactory, with coefficients of variation for these same levels of 4.73% and 4.75%, respectively. The reliability of the microalbuminuria assay results in our laboratory is confirmed by the satisfactory results obtained in our study, which comply with the recommendations of RICOS and FSCB.

GSC Biological and Pharmaceutical SciencesVol. 30(2)
Openalex Percentile: Top 96%
Chronic Kidney Disease and Diabetes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.