Evaluation of Measurement Uncertainty in Clinical Chemistry and its Comparison With Analytical Performance Specifications

Introduction To evaluate the variability in values reported by a clinical laboratory, it is important to assess the measurement uncertainty (MU). Furthermore, verifying that the estimated MU aligns with relevant analytical performance specifications (APS) is essential. Previous research has indicated variability in how laboratories estimate and apply MU, with some failing to meet APS, potentially affecting clinical decision-making. Additionally, factors such as differences in instrumentation, reagent quality, and calibration protocols may contribute to regional variations in MU, emphasizing the need for a systematic evaluation. This study aimed to determine the MU of 31 measurands and compare it with the APS available in the literature to assess whether our measuring systems can meet these APS. Methods The study was conducted in the Clinical Biochemistry Laboratory of the Himalayan Institute of Medical Sciences, Dehradun, India. The measuring systems in our laboratory are Beckman Coulter DxC 700 AU (Beckman Coulter, Inc., Brea, California, United States), Beckman Coulter UniCel DxI 800 (Beckman Coulter, Inc.,), and bioMérieux VIDAS (bioMérieux SA, Marcy-l'Étoile, France). The 'top-down' approach which uses internal quality control data and calibrator information was used for the estimation of MU. The formula used was: <i>u</i> <sub>result</sub> =√(<i>u</i> <sub>cal</sub> <sup>2</sup> + <i>u</i> <sub>Rw</sub> <sup>2</sup>+ <i>u</i> <sub>bias</sub> <sup>2</sup>). Objectively derived APS available in the literature were utilized to determine the allowable MU. Results A total of 24 measurands exhibited uncertainty estimates below the minimum APS limits. However, serum lactate dehydrogenase (LD), creatine kinase (CK), alkaline phosphatase (ALP), sodium, chloride, glucose, and ferritin showed MU higher than the minimum APS. The calibrator uncertainties (u<sub>cal</sub>) for ALP (4.7), LD (5.45), CK (11.06), and Ferritin (23.2) were significantly high, contributing to their elevated MU estimates. Additionally, the minimum APS for serum chloride (0.74) and sodium (0.40) were particularly stringent and could not be met. Conclusion The evaluation of MU provides objective insights into the quality of measurement systems and, its comparison, against set APS, supports the applicability of laboratory results in clinical decision-making. Failing to meet APS can lead to misdiagnosis, treatment errors, and patient safety risks. It may also hinder ISO 15189 accreditation and increase retesting costs.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2025-02-15
DOI
https://doi.org/10.7759/cureus.79043
Citations
1
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
1.16
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of Measurement Uncertainty in Clinical Chemistry and its Comparison With Analytical Performance Specifications

Kiesar Sideeq Bhat, Nishtha Wadhwa, M. Kalsi, Tushita Sadhu
1 citations
Cureus
Clinical Laboratory Practices and Quality Control
1.16
article

Evaluation of Measurement Uncertainty in Clinical Chemistry and its Comparison With Analytical Performance Specifications

Kiesar Sideeq Bhat, Nishtha Wadhwa, M. Kalsi, Tushita Sadhu
article en
1 citations

Abstract

Introduction To evaluate the variability in values reported by a clinical laboratory, it is important to assess the measurement uncertainty (MU). Furthermore, verifying that the estimated MU aligns with relevant analytical performance specifications (APS) is essential. Previous research has indicated variability in how laboratories estimate and apply MU, with some failing to meet APS, potentially affecting clinical decision-making. Additionally, factors such as differences in instrumentation, reagent quality, and calibration protocols may contribute to regional variations in MU, emphasizing the need for a systematic evaluation. This study aimed to determine the MU of 31 measurands and compare it with the APS available in the literature to assess whether our measuring systems can meet these APS. Methods The study was conducted in the Clinical Biochemistry Laboratory of the Himalayan Institute of Medical Sciences, Dehradun, India. The measuring systems in our laboratory are Beckman Coulter DxC 700 AU (Beckman Coulter, Inc., Brea, California, United States), Beckman Coulter UniCel DxI 800 (Beckman Coulter, Inc.,), and bioMérieux VIDAS (bioMérieux SA, Marcy-l'Étoile, France). The 'top-down' approach which uses internal quality control data and calibrator information was used for the estimation of MU. The formula used was: <i>u</i> <sub>result</sub> =√(<i>u</i> <sub>cal</sub> <sup>2</sup> + <i>u</i> <sub>Rw</sub> <sup>2</sup>+ <i>u</i> <sub>bias</sub> <sup>2</sup>). Objectively derived APS available in the literature were utilized to determine the allowable MU. Results A total of 24 measurands exhibited uncertainty estimates below the minimum APS limits. However, serum lactate dehydrogenase (LD), creatine kinase (CK), alkaline phosphatase (ALP), sodium, chloride, glucose, and ferritin showed MU higher than the minimum APS. The calibrator uncertainties (u<sub>cal</sub>) for ALP (4.7), LD (5.45), CK (11.06), and Ferritin (23.2) were significantly high, contributing to their elevated MU estimates. Additionally, the minimum APS for serum chloride (0.74) and sodium (0.40) were particularly stringent and could not be met. Conclusion The evaluation of MU provides objective insights into the quality of measurement systems and, its comparison, against set APS, supports the applicability of laboratory results in clinical decision-making. Failing to meet APS can lead to misdiagnosis, treatment errors, and patient safety risks. It may also hinder ISO 15189 accreditation and increase retesting costs.

CureusVol. 17(2)
Himalayan Institute of Yoga Science and Philosophy (US), Himalayan University (IN)
Openalex Percentile: Top 26%
Clinical Laboratory Practices and Quality Control
1.16
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.