[18F]FDG administered activity reduction capabilities of a 32-cm axial field-of-view solid-state digital bismuth germanium oxide PET/CT system while maintaining EARL compliance
Reduction in [<sup>18</sup>F]FDG administered activity is possible by at least a factor 4 for 1 min/bed with the Omni Legend 32 cm PET/CT while maintaining EARL<sub>1</sub> and EARL<sub>2</sub> compliance.
Authors
- Alina van de Burgt (ORCID: https://orcid.org/0000-0001-8170-0583)
- Lioe‐Fee de Geus‐Oei (ORCID: https://orcid.org/0000-0003-1817-2743)
- Fotis A. Kotasidis
- Daphne D. D. Rietbergen (ORCID: https://orcid.org/0000-0001-9873-6179)
- Petra Dibbets‐Schneider (ORCID: https://orcid.org/0000-0003-0538-8270)
- Floris H. P. van Velden (ORCID: https://orcid.org/0000-0003-2859-2119)
Institutions
- Leiden University Medical Center (NL)
Publication Details
- Journal
- Physica Medica
- Published
- 2025-02-15
- DOI
- https://doi.org/10.1016/j.ejmp.2025.104935
- Citations
- 3
- Primary Topic
- Medical Imaging Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 4.08
Funders
- Leids Universitair Medisch Centrum
- GE Healthcare