Inverse determination of the thermal contact conductance for an interface between a Co28Cr6Mo hip stem and a PMMA-based bone cement

Traditional mechanical methods for implant and bone cement removal during total hip arthroplasty (THA) revision surgeries typically lead to surrounding tissue damage and increased risk of femoral fractures. Transcutaneous induction heating is a promising new removal approach as it causes softening of the thermoplastic bone cement, and thus prevents damage to the surrounding tissue during removal and increases stability post-revision. However, precise knowledge of the heat transfer between implant and bone cement is necessary to minimize the risk of thermal damage to surrounding tissues. In this context, knowledge of the thermal contact conductance (TCC) at the interface of Co28Cr6Mo hip stems and PMMA-based bone cement is a key issue. The present study addresses the challenge of measuring TCC by proposing an inverse method of determination using infrared thermography measurements of the heating process and a finite element simulation with a variable parameter for the TCC. Results indicate TCC values of 3,125 ± 275 Wm<sup>- 2</sup>K<sup>- 1</sup> for dry interfaces and 5,100 ± 300 Wm<sup>- 2</sup>K<sup>- 1</sup> for wet interfaces. The influence of heat conduction on bone cement surface temperature is significant, impacting the measured surface temperatures by 15-19% for wet and 23-30% for dry interfaces. These findings are crucial for the design of heating procedures and minimization of thermal damage during induction heating assisted THA revisions.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-02-13
DOI
https://doi.org/10.1038/s41598-025-89675-w
Citations
1
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
2.01

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Inverse determination of the thermal contact conductance for an interface between a Co28Cr6Mo hip stem and a PMMA-based bone cement

Hans Jürgen Maier, Eike Jakubowitz, Sebastian Herbst, Henning Windhagen et al.
1 citations
Scientific Reports
Orthopaedic implants and arthroplasty
2.01
article

Inverse determination of the thermal contact conductance for an interface between a Co28Cr6Mo hip stem and a PMMA-based bone cement

Hans Jürgen Maier, Eike Jakubowitz, Sebastian Herbst, Henning Windhagen, Crystal Kayaro Emonde, Magnus Reulbach, Patrick Evers, Florian Nürnberger
article en
1 citations

Abstract

Traditional mechanical methods for implant and bone cement removal during total hip arthroplasty (THA) revision surgeries typically lead to surrounding tissue damage and increased risk of femoral fractures. Transcutaneous induction heating is a promising new removal approach as it causes softening of the thermoplastic bone cement, and thus prevents damage to the surrounding tissue during removal and increases stability post-revision. However, precise knowledge of the heat transfer between implant and bone cement is necessary to minimize the risk of thermal damage to surrounding tissues. In this context, knowledge of the thermal contact conductance (TCC) at the interface of Co28Cr6Mo hip stems and PMMA-based bone cement is a key issue. The present study addresses the challenge of measuring TCC by proposing an inverse method of determination using infrared thermography measurements of the heating process and a finite element simulation with a variable parameter for the TCC. Results indicate TCC values of 3,125 ± 275 Wm<sup>- 2</sup>K<sup>- 1</sup> for dry interfaces and 5,100 ± 300 Wm<sup>- 2</sup>K<sup>- 1</sup> for wet interfaces. The influence of heat conduction on bone cement surface temperature is significant, impacting the measured surface temperatures by 15-19% for wet and 23-30% for dry interfaces. These findings are crucial for the design of heating procedures and minimization of thermal damage during induction heating assisted THA revisions.

Scientific ReportsVol. 15(1)
Leibniz University Hannover (DE), Medizinische Hochschule Hannover (DE)
Deutsche Forschungsgemeinschaft, Gottfried Wilhelm Leibniz Universität Hannover
Clean water and sanitation
Openalex Percentile: Top 19%
Orthopaedic implants and arthroplasty
2.01
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.