Accumulated dose stability parameters in p-type and n-type silicon diodes

This work investigates the influence of doping type on the dose responses and the accumulated dose stability of n- and p-type silicon MCz diodes. The operating principle of diode-based dosimeters relies on measuring the radiation-induced currents delivered by non-polarized diodes throughout the exposure time. An electrometer promptly reads the current signal, linearly dose rate dependent. The offline integration of the current signal provides the charge generated in the sensitive volume of the diode, expected to be proportional to the absorbed dose. The experimental approach involves analyzing the repeatability of the current signals, the dose responses of both pristine and pre-irradiated diodes, the correspondent charge sensitivities, and the sensitivity decay with increasing doses. For doses up to 175 kGy, the results reveal a linear dose response of the MCz(p) diode, characterized by a charge sensitivity of 3.1 µC/Gy. Within the same dose range, the response of the MCz(n) diode is visibly saturated and given by a fourth-order polynomial function. This saturation effect is likely linked to radiation damage effects manifesting in the current decay with increasing accumulated doses. This surmise is confirmed in this work by a less pronounced drop in sensitivity of the p-type diode than that recorded for the n-type diode when both are subjected to 175 kGy. This behavior is ascribed to the working principle of the diode in the short-circuit current mode and the differences between the diffusion lengths of minority carriers in n- and p-type silicon materials. The diodes' response stability and dose lifespan remain to be further investigated.

Authors

Institutions

Publication Details

Journal
Brazilian Journal of Radiation Sciences
Published
2025-02-12
DOI
https://doi.org/10.15392/2319-0612.2024.2601
Primary Topic
Particle Detector Development and Performance
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accumulated dose stability parameters in p-type and n-type silicon diodes

J.A.C. Gonçalves, F. Camargo, C.C. Bueno, K. C. S. Pascoalino
Brazilian Journal of Radiation Sciences
Particle Detector Development and Performance
article

Accumulated dose stability parameters in p-type and n-type silicon diodes

J.A.C. Gonçalves, F. Camargo, C.C. Bueno, K. C. S. Pascoalino
article en

Abstract

This work investigates the influence of doping type on the dose responses and the accumulated dose stability of n- and p-type silicon MCz diodes. The operating principle of diode-based dosimeters relies on measuring the radiation-induced currents delivered by non-polarized diodes throughout the exposure time. An electrometer promptly reads the current signal, linearly dose rate dependent. The offline integration of the current signal provides the charge generated in the sensitive volume of the diode, expected to be proportional to the absorbed dose. The experimental approach involves analyzing the repeatability of the current signals, the dose responses of both pristine and pre-irradiated diodes, the correspondent charge sensitivities, and the sensitivity decay with increasing doses. For doses up to 175 kGy, the results reveal a linear dose response of the MCz(p) diode, characterized by a charge sensitivity of 3.1 µC/Gy. Within the same dose range, the response of the MCz(n) diode is visibly saturated and given by a fourth-order polynomial function. This saturation effect is likely linked to radiation damage effects manifesting in the current decay with increasing accumulated doses. This surmise is confirmed in this work by a less pronounced drop in sensitivity of the p-type diode than that recorded for the n-type diode when both are subjected to 175 kGy. This behavior is ascribed to the working principle of the diode in the short-circuit current mode and the differences between the diffusion lengths of minority carriers in n- and p-type silicon materials. The diodes' response stability and dose lifespan remain to be further investigated.

Brazilian Journal of Radiation SciencesVol. 12(4A (Suppl.))
Instituto de Pesquisas Energéticas e Nucleares (BR), Instituto de Ensinos Superiores da Amazônia (BR), National Nuclear Energy Commission (BR)
Fundação de Amparo à Pesquisa do Estado de São Paulo, Conselho Nacional de Desenvolvimento Científico e Tecnológico
Openalex Percentile: Top 98%
Particle Detector Development and Performance
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.