Efficient noble gas purification using hot getters and gas circulation by convection

Abstract Noble gas radiation detectors with optical readout are gaining popularity in fields like astrophysics and particle physics due to their ability to produce both ionization and scintillation signals in response to ionizing radiation interaction. In addition, the amplification of primary ionization signals can be achieved by promoting secondary scintillation in the gas. Noble gas purity, especially concerning impurities like H 2 O, N 2 , O 2 , CO 2 , and hydrocarbons, greatly influences its performance. These impurities can cause the loss of primary electrons and quench the scintillation signal. A very high purity level of the gas is required. In the early 90's, a simple method was developed for noble gas purification in sealed, small volume (up to few litters) gas radiation detectors. Gas purification is achieved promoting gas circulation through Zr-based hot getters, simply maintaining the gas circulation by convection. The effectiveness of this method has been only confirmed by the energy resolutions achieved in those detectors, which were similar to that achieved in other high-performance noble gas detectors. In this work, we used waveform analysis of the primary and secondary scintillation signals and we were able to evaluate the impact of the attachment and quenching caused by impurities in one of our detectors filled with pure Xe, and estimate upper values for the impurity content in the gas. The maximum overall impurity concentration was estimated to be below 4 ppm, considering nearly all the impurities, and below 82 ppm if N 2 is considered. The electron lifetime was measured to be 2.1 ± 0.1 ms, in line with those achieved in other high-performance optical detectors.

Authors

Institutions

Publication Details

Journal
Journal of Instrumentation
Published
2025-02-01
DOI
https://doi.org/10.1088/1748-0221/20/02/c02026
Citations
2
Primary Topic
Dark Matter and Cosmic Phenomena
Type
article
Field-Weighted Citation Impact
3.74
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Efficient noble gas purification using hot getters and gas circulation by convection

P. A. O. C. Silva, R. D. P. Mano, J.M.F. dos Santos, C. M. B. Monteiro et al.
2 citations
Journal of Instrumentation
Dark Matter and Cosmic Phenomena
3.74
article

Efficient noble gas purification using hot getters and gas circulation by convection

P. A. O. C. Silva, R. D. P. Mano, J.M.F. dos Santos, C. M. B. Monteiro, C. A. O. Henriques, J. M. R. Teixeira
article en
2 citations

Abstract

Abstract Noble gas radiation detectors with optical readout are gaining popularity in fields like astrophysics and particle physics due to their ability to produce both ionization and scintillation signals in response to ionizing radiation interaction. In addition, the amplification of primary ionization signals can be achieved by promoting secondary scintillation in the gas. Noble gas purity, especially concerning impurities like H 2 O, N 2 , O 2 , CO 2 , and hydrocarbons, greatly influences its performance. These impurities can cause the loss of primary electrons and quench the scintillation signal. A very high purity level of the gas is required. In the early 90's, a simple method was developed for noble gas purification in sealed, small volume (up to few litters) gas radiation detectors. Gas purification is achieved promoting gas circulation through Zr-based hot getters, simply maintaining the gas circulation by convection. The effectiveness of this method has been only confirmed by the energy resolutions achieved in those detectors, which were similar to that achieved in other high-performance noble gas detectors. In this work, we used waveform analysis of the primary and secondary scintillation signals and we were able to evaluate the impact of the attachment and quenching caused by impurities in one of our detectors filled with pure Xe, and estimate upper values for the impurity content in the gas. The maximum overall impurity concentration was estimated to be below 4 ppm, considering nearly all the impurities, and below 82 ppm if N 2 is considered. The electron lifetime was measured to be 2.1 ± 0.1 ms, in line with those achieved in other high-performance optical detectors.

Journal of InstrumentationVol. 20(02)
University of Coimbra (PT)
Openalex Percentile: Top 9%
Dark Matter and Cosmic Phenomena
3.74
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.