Accuracy of the Instantaneous Breathing and Heart Rates Estimated by Smartphone Inertial Units

Seismocardiography (SCG) and Gyrocardiography (GCG) use lightweight, miniaturized accelerometers and gyroscopes to record, respectively, cardiac-induced linear accelerations and angular velocities of the chest wall. These inertial sensors are also sensitive to thoracic movements with respiration, which cause baseline wanderings in SCG and GCG signals. Nowadays, accelerometers and gyroscopes are widely integrated into smartphones, thus increasing the potential of SCG and GCG as cardiorespiratory monitoring tools. This study investigates the accuracy of smartphone inertial sensors in simultaneously measuring instantaneous heart rates and breathing rates. Smartphone-derived SCG and GCG signals were acquired from 10 healthy subjects at rest. The performances of heartbeats and respiratory acts detection, as well as of inter-beat intervals (IBIs) and inter-breath intervals (IBrIs) estimation, were evaluated for both SCG and GCG via the comparison with simultaneous electrocardiography and respiration belt signals. Heartbeats were detected with a sensitivity and positive predictive value (PPV) of 89.3% and 93.3% in SCG signals and of 97.3% and 97.9% in GCG signals. Moreover, IBIs measurements reported strong linear relationships (R<sup>2</sup> > 0.999), non-significant biases, and Bland-Altman limits of agreement (LoA) of ±7.33 ms for SCG and ±5.22 ms for GCG. On the other hand, respiratory acts detection scored a sensitivity and PPV of 95.6% and 94.7% for SCG and of 95.7% and 92.0% for GCG. Furthermore, high R<sup>2</sup> values (0.976 and 0.968, respectively), non-significant biases, and an LoA of ±0.558 s for SCG and ±0.749 s for GCG were achieved for IBrIs estimates. The results of this study confirm that smartphone inertial sensors can provide accurate measurements of both instantaneous heart rate and breathing rate without the need for additional devices.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2025-02-12
DOI
https://doi.org/10.3390/s25041094
Citations
7
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
4.72

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accuracy of the Instantaneous Breathing and Heart Rates Estimated by Smartphone Inertial Units

Jessica Centracchio, Emilio Andreozzi, Antonio Fratini, Salvatore J. Parlato et al.
7 citations
Sensors
Non-Invasive Vital Sign Monitoring
4.72
article

Accuracy of the Instantaneous Breathing and Heart Rates Estimated by Smartphone Inertial Units

Jessica Centracchio, Emilio Andreozzi, Antonio Fratini, Salvatore J. Parlato, Daniele Esposito, Paolo Bifulco, Eliana Cinotti
article en
7 citations

Abstract

Seismocardiography (SCG) and Gyrocardiography (GCG) use lightweight, miniaturized accelerometers and gyroscopes to record, respectively, cardiac-induced linear accelerations and angular velocities of the chest wall. These inertial sensors are also sensitive to thoracic movements with respiration, which cause baseline wanderings in SCG and GCG signals. Nowadays, accelerometers and gyroscopes are widely integrated into smartphones, thus increasing the potential of SCG and GCG as cardiorespiratory monitoring tools. This study investigates the accuracy of smartphone inertial sensors in simultaneously measuring instantaneous heart rates and breathing rates. Smartphone-derived SCG and GCG signals were acquired from 10 healthy subjects at rest. The performances of heartbeats and respiratory acts detection, as well as of inter-beat intervals (IBIs) and inter-breath intervals (IBrIs) estimation, were evaluated for both SCG and GCG via the comparison with simultaneous electrocardiography and respiration belt signals. Heartbeats were detected with a sensitivity and positive predictive value (PPV) of 89.3% and 93.3% in SCG signals and of 97.3% and 97.9% in GCG signals. Moreover, IBIs measurements reported strong linear relationships (R<sup>2</sup> > 0.999), non-significant biases, and Bland-Altman limits of agreement (LoA) of ±7.33 ms for SCG and ±5.22 ms for GCG. On the other hand, respiratory acts detection scored a sensitivity and PPV of 95.6% and 94.7% for SCG and of 95.7% and 92.0% for GCG. Furthermore, high R<sup>2</sup> values (0.976 and 0.968, respectively), non-significant biases, and an LoA of ±0.558 s for SCG and ±0.749 s for GCG were achieved for IBrIs estimates. The results of this study confirm that smartphone inertial sensors can provide accurate measurements of both instantaneous heart rate and breathing rate without the need for additional devices.

SensorsVol. 25(4)
University of Salerno (IT), Aston University (GB), University of Naples Federico II (IT)
Università degli Studi di Napoli Federico II
Openalex Percentile: Top 6%
Non-Invasive Vital Sign Monitoring
4.72
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.