Patient–ventilator synchrony under non-invasive ventilation is improved by an automated real time waveform analysis algorithm: a bench study
In this NIV bench model, IntelliSync + <sup>®</sup>, used in addition to NIV-PS, improved both expiratory and inspiratory synchrony. It was particularly efficient in the presence of obstructive and restrictive respiratory mechanics and high-pressure support levels. These pre-clinical results tend to support the ability of IntelliSync + <sup>®</sup> to improve patient-ventilator synchrony in the presence of leaks and provide pre-clinical data supporting a clinical evaluation of the automated algorithm during NIV.