Test–Retest Reliability and Precision of GripAble: A Multipurpose Exergaming Device

<b><i>Objective:</i></b> GripAble is a novel upper limb motor function assessment and exergaming device. In this study, we aimed to (1) establish its test-retest reliability and precision and (2) determine which method of reporting (i.e., based on one, two, or three trials) yields the greatest reliability and precision when testing hand grip strength (HGS) in healthy-handed adults. <b><i>Materials and Methods:</i></b> Purposeful sampling was used to recruit a diverse sample of adults according to sex, age, and race. Data were collected during two sessions approximately 2 weeks apart. During each session, three trials of maximal HGS were taken bilaterally. Test-retest reliability was assessed using the intraclass correlation coefficient (ICC), and precision was assessed using the standard error of measurement and the minimal detectable change (MDC). <b><i>Results:</i></b> Thirty-six healthy-handed adults completed the study protocol. Across all methods of reporting, ICC values were ≥0.85. The highest reliability (ICC values for dominant and nondominant hands = 0.97 and 0.94, respectively) and precision (MDC% for dominant and nondominant hands = 12.9 and 18.2, respectively) were observed when the average of three trials was calculated. <b><i>Conclusions:</i></b> The GripAble device has excellent reliability and acceptable precision when used to assess maximal HGS in healthy-handed adults. Recording three trials and reporting their average yields the highest test-retest reliability and precision. <b><i>Relevance:</i></b> We present early evidence to support the use of GripAble as an HGS assessment device. Future research should include establishing HGS reference values and testing GripAble's reliability and precision in different clinical populations. Its efficacy as an exergaming intervention also warrants exploration.

Authors

Institutions

Publication Details

Journal
Games for Health Journal
Published
2025-02-12
DOI
https://doi.org/10.1089/g4h.2024.0216
Citations
1
Primary Topic
VLSI and Analog Circuit Testing
Type
article
Field-Weighted Citation Impact
2.27
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Test–Retest Reliability and Precision of GripAble: A Multipurpose Exergaming Device

Corey Weston McGee, Halil İbrahim Ergen, Roman Kudin
1 citations
Games for Health Journal
VLSI and Analog Circuit Testing
2.27
article

Test–Retest Reliability and Precision of GripAble: A Multipurpose Exergaming Device

Corey Weston McGee, Halil İbrahim Ergen, Roman Kudin
article en
1 citations

Abstract

<b><i>Objective:</i></b> GripAble is a novel upper limb motor function assessment and exergaming device. In this study, we aimed to (1) establish its test-retest reliability and precision and (2) determine which method of reporting (i.e., based on one, two, or three trials) yields the greatest reliability and precision when testing hand grip strength (HGS) in healthy-handed adults. <b><i>Materials and Methods:</i></b> Purposeful sampling was used to recruit a diverse sample of adults according to sex, age, and race. Data were collected during two sessions approximately 2 weeks apart. During each session, three trials of maximal HGS were taken bilaterally. Test-retest reliability was assessed using the intraclass correlation coefficient (ICC), and precision was assessed using the standard error of measurement and the minimal detectable change (MDC). <b><i>Results:</i></b> Thirty-six healthy-handed adults completed the study protocol. Across all methods of reporting, ICC values were ≥0.85. The highest reliability (ICC values for dominant and nondominant hands = 0.97 and 0.94, respectively) and precision (MDC% for dominant and nondominant hands = 12.9 and 18.2, respectively) were observed when the average of three trials was calculated. <b><i>Conclusions:</i></b> The GripAble device has excellent reliability and acceptable precision when used to assess maximal HGS in healthy-handed adults. Recording three trials and reporting their average yields the highest test-retest reliability and precision. <b><i>Relevance:</i></b> We present early evidence to support the use of GripAble as an HGS assessment device. Future research should include establishing HGS reference values and testing GripAble's reliability and precision in different clinical populations. Its efficacy as an exergaming intervention also warrants exploration.

Games for Health JournalVol. 14(4)
University of Minnesota (US), University of Minnesota Medical Center (US), Gaziantep University (TR)
Openalex Percentile: Top 18%
VLSI and Analog Circuit Testing
2.27
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.