Comparative accuracy of Plusoptix A16, Vision Screener V100, and Grand Seiko WAM-5500 autorefractors versus subjective refraction
Significant differences in refractive error measurements were observed between devices, with the Vision Screener V100 overestimating errors and the Grand Seiko WAM-5500 showing higher J0 values. Subjective refraction detected more with-the-rule astigmatism, highlighting the importance of validating autorefraction results, especially in high astigmatism cases. Age-related differences in M and J0 components highlight the need to confirm objective methods with subjective refraction.
Authors
- Jorge Jorge (ORCID: https://orcid.org/0000-0002-4369-3332)
- Paulo Fernandes (ORCID: https://orcid.org/0000-0002-3921-6357)
Institutions
- University of Minho (PT)
Publication Details
- Journal
- Clinical and Experimental Optometry
- Published
- 2025-02-12
- DOI
- https://doi.org/10.1080/08164622.2025.2458161
- Primary Topic
- Ophthalmology and Visual Impairment Studies
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Foundation for Science and Technology