Spectral performance for iodine quantification of a dual‐source, dual‐kV photon counting detector CT

SS-PCD-CT with two energy thresholds outperformed DS-EID-CT in terms of image noise in iodine basis images when spatial resolution was matched. DS-PCD-CT gave the lowest noise due to the combination of PCD technology and improved spectral separation from the different x-ray spectra.

Authors

Institutions

Publication Details

Journal
Medical Physics
Published
2025-02-10
DOI
https://doi.org/10.1002/mp.17679
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Spectral performance for iodine quantification of a dual‐source, dual‐kV photon counting detector CT

Kevin Treb, Cynthia H. McCollough, Tim N. Winfree, Shuai Leng
Medical Physics
Advanced X-ray and CT Imaging
article

Spectral performance for iodine quantification of a dual‐source, dual‐kV photon counting detector CT

Kevin Treb, Cynthia H. McCollough, Tim N. Winfree, Shuai Leng
article en

Abstract

SS-PCD-CT with two energy thresholds outperformed DS-EID-CT in terms of image noise in iodine basis images when spatial resolution was matched. DS-PCD-CT gave the lowest noise due to the combination of PCD technology and improved spectral separation from the different x-ray spectra.

Medical PhysicsVol. 52(5)
Mayo Clinic (US)
National Institutes of Health
Affordable and clean energy
Openalex Percentile: Top 98%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.