CUSTOM 3D-PRINTED IMPLANTS FOR CRITICAL-SIZED BONE DEFECTS IN FOOT AND ANKLE SURGERY: A MULTICENTRE COLLABORATIVE STUDY

Introduction The treatment of critical-sized bone defects in foot and ankle surgery remains challenging. Traditional methods, such as bone transport, bulk allograft, vascularized bone graft, and Masquelet procedures carry risks including multiple surgeries, donor site morbidity, infection, and non-union. Recently, custom 3D-printed implants have emerged, offering improved anatomical compatibility, eliminating the need for tissue harvesting, and often requiring only a single operation. This study aims to present the largest UK series of custom 3D-printed implants in foot and ankle surgery. Methods A retrospective multicentre collaborative study was conducted in 10 NHS Trusts. Demographic, surgical and radiographic variables were recorded including: age; sex; side; type of operation; complications; and postop imaging at 6 weeks, 3 months, 6 months and 12 months. Baseline characteristics were described for demographic variables. Categorical variables were expressed as frequencies and percentages. Results Data was collected on a total of 34 patients. The mean age of patients was 54 years (range 29 to 83). 68% of patients were men. Avascular necrosis was the most common indication for surgery (35%) followed by trauma (21%), total ankle replacement failure (21%), infection (12%), Charcot arthropathy (6%) and AVN and infection (3%). The most common operation performed was a fusion augmented with a truss (44%) and 91% of patients had a combination of regional and general anaesthetic. All patients had an ASA grade of less than 4. Of the patients that had follow-up imaging, the implants were reported intact by 94% at 6 weeks, 78% at 3 months, 68% at 6 months and 67% at 12 months. Discussion This is the largest case series of custom-made foot and ankle implants in the UK to date. The clinical and radiographic outcomes at one year Follow-up are satisfactory. Further work is required to assess long term implant survivorship.

Authors

Institutions

Publication Details

Journal
Orthopaedic Proceedings
Published
2025-02-10
DOI
https://doi.org/10.1302/1358-992x.2025.1.010
Primary Topic
Anatomy and Medical Technology
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

CUSTOM 3D-PRINTED IMPLANTS FOR CRITICAL-SIZED BONE DEFECTS IN FOOT AND ANKLE SURGERY: A MULTICENTRE COLLABORATIVE STUDY

Linzy Houchen-Wolloff, Jitendra Mangwani, Abbas See
Orthopaedic Proceedings
Anatomy and Medical Technology
article

CUSTOM 3D-PRINTED IMPLANTS FOR CRITICAL-SIZED BONE DEFECTS IN FOOT AND ANKLE SURGERY: A MULTICENTRE COLLABORATIVE STUDY

Linzy Houchen-Wolloff, Jitendra Mangwani, Abbas See
article en

Abstract

Introduction The treatment of critical-sized bone defects in foot and ankle surgery remains challenging. Traditional methods, such as bone transport, bulk allograft, vascularized bone graft, and Masquelet procedures carry risks including multiple surgeries, donor site morbidity, infection, and non-union. Recently, custom 3D-printed implants have emerged, offering improved anatomical compatibility, eliminating the need for tissue harvesting, and often requiring only a single operation. This study aims to present the largest UK series of custom 3D-printed implants in foot and ankle surgery. Methods A retrospective multicentre collaborative study was conducted in 10 NHS Trusts. Demographic, surgical and radiographic variables were recorded including: age; sex; side; type of operation; complications; and postop imaging at 6 weeks, 3 months, 6 months and 12 months. Baseline characteristics were described for demographic variables. Categorical variables were expressed as frequencies and percentages. Results Data was collected on a total of 34 patients. The mean age of patients was 54 years (range 29 to 83). 68% of patients were men. Avascular necrosis was the most common indication for surgery (35%) followed by trauma (21%), total ankle replacement failure (21%), infection (12%), Charcot arthropathy (6%) and AVN and infection (3%). The most common operation performed was a fusion augmented with a truss (44%) and 91% of patients had a combination of regional and general anaesthetic. All patients had an ASA grade of less than 4. Of the patients that had follow-up imaging, the implants were reported intact by 94% at 6 weeks, 78% at 3 months, 68% at 6 months and 67% at 12 months. Discussion This is the largest case series of custom-made foot and ankle implants in the UK to date. The clinical and radiographic outcomes at one year Follow-up are satisfactory. Further work is required to assess long term implant survivorship.

Orthopaedic ProceedingsVol. 107-B(SUPP_1)
University of Nottingham (GB), University Hospitals of Leicester NHS Trust (GB)
Openalex Percentile: Top 98%
Anatomy and Medical Technology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.