Corneal Imaging and Densitometry Measurements in Patients with Fuchs' Dystrophy Undergoing Penetrating Keratoplasty and Descemet's Striping Automated Endothelial Keratoplasty

Aims: In the present study, we used the densitometry software from the Oculus Pentacam to compare postoperative corneal clarity between penetrating keratoplasty (PK) and Descemet’s stripping endothelial keratoplasty (DSAEK) in patients with Fuchs’ dystrophy. Methods: A retrospective comparative study was carried out at Manchester Royal Eye Hospital. In 28 patients with Fuchs’ dystrophy, corneal densitometry measurements were performed 12–18 months after corneal transplantation. The correlations of the densitometry measurements with the best corrected visual acuity (BCVA) and central corneal thickness (CCT) were analysed and compared between eyes that underwent PK and those that underwent DSAEK. Results: Corneal densitometry measurements in the 33 eyes showed no significant differences between the PK and DSAEK post-surgery groups. There was no significant correlation between CCT and corneal densitometry measurements in either group (P >0.05 in both cases). After DSAEK, corneal densitometry measurements were significantly correlated with BCVA in the central (P = 0.01), posterior (P = 0.007), and full-depth (P = 0.008) 0–2 mm zones of the cornea but not in PK group. The postoperative CCT was significantly different between the two groups (P <0.01). Conclusion: The 2 types of corneal transplantation resulted in different outcomes in terms of corneal densitometry measurements. Improved visual acuity after DSAEK was found to positively correlate with improvements in corneal clarity. Oculus Pentacam provides an objective evaluation tool to monitor corneal status after surgery.

Authors

Institutions

Publication Details

Journal
OphthaTherapy Therapies in Ophthalmology
Published
2024-12-31
DOI
https://doi.org/10.24292/01.ot.212124
Primary Topic
Corneal surgery and disorders
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Corneal Imaging and Densitometry Measurements in Patients with Fuchs' Dystrophy Undergoing Penetrating Keratoplasty and Descemet's Striping Automated Endothelial Keratoplasty

M Chantal Hillarby, Jawaher Alshehri, Khaled Alzahrani, Fiona Carley et al.
OphthaTherapy Therapies in Ophthalmology
Corneal surgery and disorders
article

Corneal Imaging and Densitometry Measurements in Patients with Fuchs' Dystrophy Undergoing Penetrating Keratoplasty and Descemet's Striping Automated Endothelial Keratoplasty

M Chantal Hillarby, Jawaher Alshehri, Khaled Alzahrani, Fiona Carley, Arun Brahma
article en

Abstract

Aims: In the present study, we used the densitometry software from the Oculus Pentacam to compare postoperative corneal clarity between penetrating keratoplasty (PK) and Descemet’s stripping endothelial keratoplasty (DSAEK) in patients with Fuchs’ dystrophy. Methods: A retrospective comparative study was carried out at Manchester Royal Eye Hospital. In 28 patients with Fuchs’ dystrophy, corneal densitometry measurements were performed 12–18 months after corneal transplantation. The correlations of the densitometry measurements with the best corrected visual acuity (BCVA) and central corneal thickness (CCT) were analysed and compared between eyes that underwent PK and those that underwent DSAEK. Results: Corneal densitometry measurements in the 33 eyes showed no significant differences between the PK and DSAEK post-surgery groups. There was no significant correlation between CCT and corneal densitometry measurements in either group (P >0.05 in both cases). After DSAEK, corneal densitometry measurements were significantly correlated with BCVA in the central (P = 0.01), posterior (P = 0.007), and full-depth (P = 0.008) 0–2 mm zones of the cornea but not in PK group. The postoperative CCT was significantly different between the two groups (P <0.01). Conclusion: The 2 types of corneal transplantation resulted in different outcomes in terms of corneal densitometry measurements. Improved visual acuity after DSAEK was found to positively correlate with improvements in corneal clarity. Oculus Pentacam provides an objective evaluation tool to monitor corneal status after surgery.

OphthaTherapy Therapies in OphthalmologyVol. 11(4)
Manchester Royal Eye Hospital (GB), Al Baha University (SA)
University of Manchester
Openalex Percentile: Top 57%
Corneal surgery and disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.