Comparison of Three Quality Assurance Methods for Clinical Ultrasound Devices

Abstract Purpose The purpose of the work was to compare the performance of different ultrasound quality assurance methods in finding transducer defects. For this, three different commonly used ultrasound QA methods were studied and compared. Methods The integrity of 99 US transducers was measured using the FirstCall transducer test in two tertiary healthcare centers in Finland. Thereafter, the ability of three different quality assurance methods (the FirstCall sensitivity test, in-air reverberation test, and uniformity test) to find defective probes was evaluated with a sub-group (N=37) of probes. All sensors were classified as either defective or functional according to predefined criteria. Results According to the 99 FirstCall measurement reports, 31% of the transducers were defective. The sensitivity test, in-air test, and uniformity test identified 27, 27, and 14% of the transducers, respectively, as defective. The agreement between the three quality assurance methods was 78%. Conclusion Even if ultrasound quality assurance is performed systematically, the number of undetected defective transducers can still be high. FirstCall allows the transducers to be measured independently and is useful in testing the sensitivity and capacitance of each element and the wiring. The in-air reverberation test is effective in finding transducer failures. It is the easiest and fastest to implement and is therefore recommended to be performed regularly for all ultrasound transducers in clinical use along with the mechanical inspection. The uniformity test requires a phantom and can be used to confirm a possible defect found by the two other methods.

Authors

Institutions

Publication Details

Journal
Journal of Medical and Biological Engineering
Published
2025-02-01
DOI
https://doi.org/10.1007/s40846-025-00932-9
Citations
2
Primary Topic
Flow Measurement and Analysis
Type
article
Field-Weighted Citation Impact
1.73
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of Three Quality Assurance Methods for Clinical Ultrasound Devices

Anna Vuorenmaa, Eetu Siitama, Pia Linder, Anastasia Koivikko et al.
2 citations
Journal of Medical and Biological Engineering
Flow Measurement and Analysis
1.73
article

Comparison of Three Quality Assurance Methods for Clinical Ultrasound Devices

Anna Vuorenmaa, Eetu Siitama, Pia Linder, Anastasia Koivikko, Hannu Eskola, J. Aarnio, Jari Heikkinen
article en
2 citations

Abstract

Abstract Purpose The purpose of the work was to compare the performance of different ultrasound quality assurance methods in finding transducer defects. For this, three different commonly used ultrasound QA methods were studied and compared. Methods The integrity of 99 US transducers was measured using the FirstCall transducer test in two tertiary healthcare centers in Finland. Thereafter, the ability of three different quality assurance methods (the FirstCall sensitivity test, in-air reverberation test, and uniformity test) to find defective probes was evaluated with a sub-group (N=37) of probes. All sensors were classified as either defective or functional according to predefined criteria. Results According to the 99 FirstCall measurement reports, 31% of the transducers were defective. The sensitivity test, in-air test, and uniformity test identified 27, 27, and 14% of the transducers, respectively, as defective. The agreement between the three quality assurance methods was 78%. Conclusion Even if ultrasound quality assurance is performed systematically, the number of undetected defective transducers can still be high. FirstCall allows the transducers to be measured independently and is useful in testing the sensitivity and capacitance of each element and the wiring. The in-air reverberation test is effective in finding transducer failures. It is the easiest and fastest to implement and is therefore recommended to be performed regularly for all ultrasound transducers in clinical use along with the mechanical inspection. The uniformity test requires a phantom and can be used to confirm a possible defect found by the two other methods.

Journal of Medical and Biological EngineeringVol. 45(1)
Tampere University (FI), Kuopio University Hospital (FI), Pirkanmaa Hospital District (FI), Tampere University Hospital (FI), Tampere University (FI)
Openalex Percentile: Top 20%
Flow Measurement and Analysis
1.73
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.