High-Performance Dual-Ended SiPM Readout for TOF-PET With BGO and LYSO:Ce

Detection time performance is a key aspect for time-of-flight positron emission tomography. With recent advancement in SiPM technology and fast readout electronics, one limiting factor on timing performance is light transport in the crystal. For high aspect-ratio crystals with single-ended readout, the time information of approximately half the optical photons is severely degraded as they initially travel in the direction opposed to the photodetector. For promptly-emitted Cherenkov photons, the increase of variance of optical path length limits their intrinsic advantage. Low-noise and high-frequency dual-ended SiPM readout can be employed to mitigate the aforementioned challenges and has the potential to combine ultra-fast timing with highest gamma-ray detection efficiency. We have studied the timing properties of cerium-doped lutetium-yttrium-oxyorthosilicate (LYSO:Ce) and bismuth germanate (BGO) in a symmetric dual-ended SiPM readout configuration. A time-based depth-of-interaction correction and a novel adaptive timestamp weighting was used to optimize the timing performance. Coupling 3×3×20 mm<sup>3</sup> polished BGO crystals to Broadcom AFBR-S4N44P014M SiPMs a CTR of 234 ± 4 ps FWHM (harmonic average) was obtained for all photopeak events. For same-sized LYSO:Ce crystals, the measured CTR value is 98 ± 2 ps, which is in excellent agreement with analytic calculations on the timing limits considering scintillation properties and modeling of light transport. The results demonstrate significant timing improvement with dual-ended readout, both for Cherenkov photons in BGO and for standard scintillation for enhanced diagnostic accuracy in PET imaging.

Authors

Institutions

Publication Details

Journal
IEEE Transactions on Radiation and Plasma Medical Sciences
Published
2025-02-07
DOI
https://doi.org/10.1109/trpms.2025.3539191
Citations
7
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
8.16

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

High-Performance Dual-Ended SiPM Readout for TOF-PET With BGO and LYSO:Ce

Emilie Roncali, Gerard Ariño‐Estrada, N. Kratochwil, Joshua W. Cates
7 citations
IEEE Transactions on Radiation and Plasma Medical Sciences
Medical Imaging Techniques and Applications
8.16
article

High-Performance Dual-Ended SiPM Readout for TOF-PET With BGO and LYSO:Ce

Emilie Roncali, Gerard Ariño‐Estrada, N. Kratochwil, Joshua W. Cates
article en
7 citations

Abstract

Detection time performance is a key aspect for time-of-flight positron emission tomography. With recent advancement in SiPM technology and fast readout electronics, one limiting factor on timing performance is light transport in the crystal. For high aspect-ratio crystals with single-ended readout, the time information of approximately half the optical photons is severely degraded as they initially travel in the direction opposed to the photodetector. For promptly-emitted Cherenkov photons, the increase of variance of optical path length limits their intrinsic advantage. Low-noise and high-frequency dual-ended SiPM readout can be employed to mitigate the aforementioned challenges and has the potential to combine ultra-fast timing with highest gamma-ray detection efficiency. We have studied the timing properties of cerium-doped lutetium-yttrium-oxyorthosilicate (LYSO:Ce) and bismuth germanate (BGO) in a symmetric dual-ended SiPM readout configuration. A time-based depth-of-interaction correction and a novel adaptive timestamp weighting was used to optimize the timing performance. Coupling 3×3×20 mm<sup>3</sup> polished BGO crystals to Broadcom AFBR-S4N44P014M SiPMs a CTR of 234 ± 4 ps FWHM (harmonic average) was obtained for all photopeak events. For same-sized LYSO:Ce crystals, the measured CTR value is 98 ± 2 ps, which is in excellent agreement with analytic calculations on the timing limits considering scintillation properties and modeling of light transport. The results demonstrate significant timing improvement with dual-ended readout, both for Cherenkov photons in BGO and for standard scintillation for enhanced diagnostic accuracy in PET imaging.

IEEE Transactions on Radiation and Plasma Medical SciencesVol. 9(6)
Lawrence Berkeley National Laboratory (US), University of California, Davis (US)
National Institute for Nutrition and Health
Openalex Percentile: Top 3%
Medical Imaging Techniques and Applications
8.16
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.