Transient temperature measurement of electron beam bombarded stainless steel via nanosecond two-color pyrometry

A two-color pyrometer is presented that enables detection of a temperature rise with temporal resolution of 17.5 ns. It is geared toward extreme environments where fast electrical electron pulses or currents may push the temperature of a metal surface to melt on a 100 ns time scale. The developed pyrometer, which utilizes an internal gain of 100x, has been tested for SS304 surfaces subjected to an intense electron beam in vacuum. The two-color pyrometer is calibrated using a quasi-calibration method by heating a graphite disk and measuring this temperature with a FLIR TG297 thermal camera. The quasi-calibration method supports the pyrometer to be within 3% accuracy of absolute temperature. The measured temporal temperature profile of the SS304 surface was derived from the pyrometer output data through a detailed gray-body analysis utilizing empirical emissivity data from publicly available literature.

Authors

Institutions

Publication Details

Journal
Review of Scientific Instruments
Published
2025-02-01
DOI
https://doi.org/10.1063/5.0215582
Primary Topic
Plasma Diagnostics and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Transient temperature measurement of electron beam bombarded stainless steel via nanosecond two-color pyrometry

J. Dickens, J. Mockert, A. Neuber, Taylor K. Watson
Review of Scientific Instruments
Plasma Diagnostics and Applications
article

Transient temperature measurement of electron beam bombarded stainless steel via nanosecond two-color pyrometry

J. Dickens, J. Mockert, A. Neuber, Taylor K. Watson
article en

Abstract

A two-color pyrometer is presented that enables detection of a temperature rise with temporal resolution of 17.5 ns. It is geared toward extreme environments where fast electrical electron pulses or currents may push the temperature of a metal surface to melt on a 100 ns time scale. The developed pyrometer, which utilizes an internal gain of 100x, has been tested for SS304 surfaces subjected to an intense electron beam in vacuum. The two-color pyrometer is calibrated using a quasi-calibration method by heating a graphite disk and measuring this temperature with a FLIR TG297 thermal camera. The quasi-calibration method supports the pyrometer to be within 3% accuracy of absolute temperature. The measured temporal temperature profile of the SS304 surface was derived from the pyrometer output data through a detailed gray-body analysis utilizing empirical emissivity data from publicly available literature.

Review of Scientific InstrumentsVol. 96(2)
Texas Tech University (US)
Affordable and clean energy
Openalex Percentile: Top 98%
Plasma Diagnostics and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.