Agreement of Automated Oscillometric Blood Pressure Measurement Device with the Manual Mercury Sphygmomanometer

The usage of the manual mercury sphygmomanometer to measure blood pressure (BP) is universally regarded as the gold standard, but its use has decreased globally due to concerns about potential mercury toxicity. Automated BP devices have become more popular, though their accuracy in detecting hypertension hasn't been studied extensively. This cross-sectional study was conducted on 301 adult female students of the University of Delhi with the aim to assess the degree of agreement between standard manual mercury sphygmomanometer and automated BP monitor (Omron JPN1). Three consecutive BP readings were taken on the left arm by a certified medical doctor using both instruments. Bland Altman plots were created to assess the degree of agreement between the two BP devices and to test the accuracy of automated BP monitor in detecting hypertensive adults. The automated BP monitor underestimated the systolic BP measurement by 4 mmHg and diastolic BP measurement by 7 mmHg as determined by Bland Altman plot. Based on international validation protocols, a mean difference of < 5 ± 8 mmHg was established to assess the level of agreement between the two devices. The automatic BP monitor did not show complete agreement with the "gold standard" mercury sphygmomanometer. With growing restrictions on the use of mercury sphygmomanometers, its replacement with an equivalent mercury-free validated device is urgently needed.

Authors

Institutions

Publication Details

Journal
Indian Journal of Community Medicine
Published
2025-02-06
DOI
https://doi.org/10.4103/ijcm.ijcm_590_23
Primary Topic
Heart Rate Variability and Autonomic Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Agreement of Automated Oscillometric Blood Pressure Measurement Device with the Manual Mercury Sphygmomanometer

Sana Ahmed, R. P. Thakur, Geeta Trilok‐Kumar, Renuka Pathak et al.
Indian Journal of Community Medicine
Heart Rate Variability and Autonomic Control
article

Agreement of Automated Oscillometric Blood Pressure Measurement Device with the Manual Mercury Sphygmomanometer

Sana Ahmed, R. P. Thakur, Geeta Trilok‐Kumar, Renuka Pathak, Yamini Gusain
article en

Abstract

The usage of the manual mercury sphygmomanometer to measure blood pressure (BP) is universally regarded as the gold standard, but its use has decreased globally due to concerns about potential mercury toxicity. Automated BP devices have become more popular, though their accuracy in detecting hypertension hasn't been studied extensively. This cross-sectional study was conducted on 301 adult female students of the University of Delhi with the aim to assess the degree of agreement between standard manual mercury sphygmomanometer and automated BP monitor (Omron JPN1). Three consecutive BP readings were taken on the left arm by a certified medical doctor using both instruments. Bland Altman plots were created to assess the degree of agreement between the two BP devices and to test the accuracy of automated BP monitor in detecting hypertensive adults. The automated BP monitor underestimated the systolic BP measurement by 4 mmHg and diastolic BP measurement by 7 mmHg as determined by Bland Altman plot. Based on international validation protocols, a mean difference of < 5 ± 8 mmHg was established to assess the level of agreement between the two devices. The automatic BP monitor did not show complete agreement with the "gold standard" mercury sphygmomanometer. With growing restrictions on the use of mercury sphygmomanometers, its replacement with an equivalent mercury-free validated device is urgently needed.

Indian Journal of Community MedicineVol. 50(4)
University of Delhi (IN), Ashoka University (IN)
Openalex Percentile: Top 96%
Heart Rate Variability and Autonomic Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.