Quality in Question: Assessing the Accuracy of Four Heart Rate Wearables and the Implications for Psychophysiological Research

Heart rate (HR) and heart rate variability (HRV) are two key measures with significant relevance in psychophysiological studies, and their measurement has become more convenient due to advances in wearable technology. However, photoplethysmography (PPG)-based wearables pose critical validity concerns. In this study, we validated four PPG wearables: three consumer-grade devices (Kyto2935, Schone Rhythm 24, and HeartMath Inner Balance Bluetooth) and one research-grade device (Empatica EmbracePlus, successor to the widely-used but discontinued Empatica E4). All devices were worn simultaneously by 40 healthy participants who underwent conditions commonly used in laboratory research (seated rest, arithmetic task, recovery, slow-paced breathing, a neuropsychological task, posture manipulation by standing up) and encountered in ambulatory-like settings (slow walking and stationary biking), compared against a criterion electrocardiography device, the Vrije Universiteit Ambulatory Monitoring System (VU-AMS). We determined the signal quality, the linear strength through regression analysis, the bias through Bland-Altman analysis, and the measurement error through mean arctangent absolute percentage error for each condition against the criterion device. We found that the research-grade device did not outperform the consumer-grade devices in laboratory conditions. It also showed low agreement with the ECG in ambulatory-like conditions. In general, devices captured HR more accurately than HRV. Finally, conditions that deviated from baseline settings and involved slight to high movement, negatively impacted the agreement between PPG devices and the criterion. We conclude that PPG devices, even those advertised and designed for research purposes, may pose validity concerns for HRV measurement in conditions other than those similar to resting states.

Authors

Institutions

Publication Details

Journal
Psychophysiology
Published
2025-02-01
DOI
https://doi.org/10.1111/psyp.70004
Citations
5
Primary Topic
Heart Rate Variability and Autonomic Control
Type
article
Field-Weighted Citation Impact
6.94

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quality in Question: Assessing the Accuracy of Four Heart Rate Wearables and the Implications for Psychophysiological Research

Martin Gevonden, Lydia Krabbendam, Mohammadamin Sinichi
5 citations
Psychophysiology
Heart Rate Variability and Autonomic Control
6.94
article

Quality in Question: Assessing the Accuracy of Four Heart Rate Wearables and the Implications for Psychophysiological Research

Martin Gevonden, Lydia Krabbendam, Mohammadamin Sinichi
article en
5 citations

Abstract

Heart rate (HR) and heart rate variability (HRV) are two key measures with significant relevance in psychophysiological studies, and their measurement has become more convenient due to advances in wearable technology. However, photoplethysmography (PPG)-based wearables pose critical validity concerns. In this study, we validated four PPG wearables: three consumer-grade devices (Kyto2935, Schone Rhythm 24, and HeartMath Inner Balance Bluetooth) and one research-grade device (Empatica EmbracePlus, successor to the widely-used but discontinued Empatica E4). All devices were worn simultaneously by 40 healthy participants who underwent conditions commonly used in laboratory research (seated rest, arithmetic task, recovery, slow-paced breathing, a neuropsychological task, posture manipulation by standing up) and encountered in ambulatory-like settings (slow walking and stationary biking), compared against a criterion electrocardiography device, the Vrije Universiteit Ambulatory Monitoring System (VU-AMS). We determined the signal quality, the linear strength through regression analysis, the bias through Bland-Altman analysis, and the measurement error through mean arctangent absolute percentage error for each condition against the criterion device. We found that the research-grade device did not outperform the consumer-grade devices in laboratory conditions. It also showed low agreement with the ECG in ambulatory-like conditions. In general, devices captured HR more accurately than HRV. Finally, conditions that deviated from baseline settings and involved slight to high movement, negatively impacted the agreement between PPG devices and the criterion. We conclude that PPG devices, even those advertised and designed for research purposes, may pose validity concerns for HRV measurement in conditions other than those similar to resting states.

PsychophysiologyVol. 62(2)
Vrije Universiteit Amsterdam (NL)
Wellcome Trust
Openalex Percentile: Top 3%
Heart Rate Variability and Autonomic Control
6.94
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.