Determination of Radiation Dose Rate and Dose Analysis of Radiation Workers in Radiology Installation at Pariaman Hospital

Diagnostic radiology is the most significant contributor to irradiation in the general population. Unnecessary irradiation of workers, patients, and the public during radiological procedures should be significantly reduced without affecting the medical diagnostic value. This study was conducted to analyze and evaluate the radiation dose rate, radiation protection facilities and radiation worker dose of Radiology Installation of Pariaman Regional Hospital. The method of determining the radiation dose rate at the Radiology Instalansi of Pariaman Hospital uses a PDR 303 gamma surveymeter and the rooms to be measured are CT-Scan rooms, conventional X-rays and mammography. Collecting data on radiation protection facilities, then it will be analyzed based on Nuclear Energy Regulatory Agency Regulation No. 4 of 2020 and radiation worker dose data in 2023 will be analyzed based on Nuclear Energy Regulatory Agency Regulation No. 4 of 2013. The results of the study showed that the radiation dose rate around the CT-Scan room was found to be (0.24-1.61) μSv/hour, the conventional X-ray room with irradiation using the minimum exposure factor obtained a radiation dose rate around the room ranging from (0.013-0.33) μSv/hour and with the maximum exposure factor obtained a dose rate around the room of (0.33-1.08) μSv/hour and the mammography room obtained a radiation dose rate of (0.03-0.40) μSv/hour. The radiology room facilities used by Pariaman Hospital are in accordance with the standards set by Nuclear Energy Regulatory Agency Regulation No. 4 of 2020. And the dose rate received by radiation workers at Pariaman Hospital is below the Dose Limit Value set by Nuclear Energy Regulatory Agency Regulation No. 4 of 2013. Based on the results of this study, it is found that there are several measurement points in the CT-Scan and conventional X-ray rooms that exceed the community dose limit value.

Authors

Institutions

Publication Details

Journal
Jurnal Fisika Flux Jurnal Ilmiah Fisika FMIPA Universitas Lambung Mangkurat
Published
2025-01-22
DOI
https://doi.org/10.20527/flux.v21i3.20529
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Determination of Radiation Dose Rate and Dose Analysis of Radiation Workers in Radiology Installation at Pariaman Hospital

Ramacos Fardela, Feriska Handayani Irka, Almahdi Mousa, Susi Andriati et al.
Jurnal Fisika Flux Jurnal Ilmiah Fisika FMIPA Universitas Lambung Mangkurat
Radiation Dose and Imaging
article

Determination of Radiation Dose Rate and Dose Analysis of Radiation Workers in Radiology Installation at Pariaman Hospital

Ramacos Fardela, Feriska Handayani Irka, Almahdi Mousa, Susi Andriati, Ade Wahyuni
article en

Abstract

Diagnostic radiology is the most significant contributor to irradiation in the general population. Unnecessary irradiation of workers, patients, and the public during radiological procedures should be significantly reduced without affecting the medical diagnostic value. This study was conducted to analyze and evaluate the radiation dose rate, radiation protection facilities and radiation worker dose of Radiology Installation of Pariaman Regional Hospital. The method of determining the radiation dose rate at the Radiology Instalansi of Pariaman Hospital uses a PDR 303 gamma surveymeter and the rooms to be measured are CT-Scan rooms, conventional X-rays and mammography. Collecting data on radiation protection facilities, then it will be analyzed based on Nuclear Energy Regulatory Agency Regulation No. 4 of 2020 and radiation worker dose data in 2023 will be analyzed based on Nuclear Energy Regulatory Agency Regulation No. 4 of 2013. The results of the study showed that the radiation dose rate around the CT-Scan room was found to be (0.24-1.61) μSv/hour, the conventional X-ray room with irradiation using the minimum exposure factor obtained a radiation dose rate around the room ranging from (0.013-0.33) μSv/hour and with the maximum exposure factor obtained a dose rate around the room of (0.33-1.08) μSv/hour and the mammography room obtained a radiation dose rate of (0.03-0.40) μSv/hour. The radiology room facilities used by Pariaman Hospital are in accordance with the standards set by Nuclear Energy Regulatory Agency Regulation No. 4 of 2020. And the dose rate received by radiation workers at Pariaman Hospital is below the Dose Limit Value set by Nuclear Energy Regulatory Agency Regulation No. 4 of 2013. Based on the results of this study, it is found that there are several measurement points in the CT-Scan and conventional X-ray rooms that exceed the community dose limit value.

Jurnal Fisika Flux Jurnal Ilmiah Fisika FMIPA Universitas Lambung MangkuratVol. 21(3)
Al-Arab Medical University (LY), University Hospitals Parma Medical Center (US), Andalas University (ID)
Openalex Percentile: Top 97%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.