Discordance between 90Y-PET/CT(MR)-estimated activity and dose calibrator measured administered activity: an international study in SIRT patients treated with resin and glass microspheres
We observed in patients similar activity discrepancies as reported for vials, with a relative difference of 44 ± 16% between glass and resin <sup>90</sup>Y-loaded microspheres. In <sup>90</sup>Y hepatic radioembolization, the 10% accuracy prerequisite on knowing the administered therapeutic activity is then unlikely to be met.