Evaluation of Timepix3 with a 1 mm thick silicon sensor as a Compton imaging polarimeter in the hard X-ray band

Abstract The present work demonstrates the usage of a Timepix3 with a 1 mm thick silicon sensor as a scattering polarimeter with simultaneous Compton-camera imaging. For this purpose, laboratory experiments were designed where the detector was placed at different scattering angles with respect to a plastic target that is irradiated with X-rays from a Hamamatsu microfocus tube (tube voltage 75 kV, tube current 60 µA). Photons scattered in the target at 90 ∘ are almost 100 % linearly polarized. Measurements in an energy range of 32.5–67.5 keV showed modulation factors above 70 %. Simulations of 100 % linearly polarized beams showed modulation factors μ 100 > 77 % in an energy range of 45–80 keV. The efficiency of detection has a maximum above 0.13 % at 45–50 keV. We implemented the origin ensemble with resolution recovery (OE-RR) algorithm for Compton camera imaging. The full width at half maximum (FWHM) of the image of 16–21 ∘ was achieved.

Authors

Institutions

Publication Details

Journal
Journal of Instrumentation
Published
2025-02-01
DOI
https://doi.org/10.1088/1748-0221/20/02/c02006
Citations
2
Primary Topic
Particle Detector Development and Performance
Type
article
Field-Weighted Citation Impact
3.74
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of Timepix3 with a 1 mm thick silicon sensor as a Compton imaging polarimeter in the hard X-ray band

Jindřich Jelínek, Petr Smolyanskiy, B. Bergmann
2 citations
Journal of Instrumentation
Particle Detector Development and Performance
3.74
article

Evaluation of Timepix3 with a 1 mm thick silicon sensor as a Compton imaging polarimeter in the hard X-ray band

Jindřich Jelínek, Petr Smolyanskiy, B. Bergmann
article en
2 citations

Abstract

Abstract The present work demonstrates the usage of a Timepix3 with a 1 mm thick silicon sensor as a scattering polarimeter with simultaneous Compton-camera imaging. For this purpose, laboratory experiments were designed where the detector was placed at different scattering angles with respect to a plastic target that is irradiated with X-rays from a Hamamatsu microfocus tube (tube voltage 75 kV, tube current 60 µA). Photons scattered in the target at 90 ∘ are almost 100 % linearly polarized. Measurements in an energy range of 32.5–67.5 keV showed modulation factors above 70 %. Simulations of 100 % linearly polarized beams showed modulation factors μ 100 > 77 % in an energy range of 45–80 keV. The efficiency of detection has a maximum above 0.13 % at 45–50 keV. We implemented the origin ensemble with resolution recovery (OE-RR) algorithm for Compton camera imaging. The full width at half maximum (FWHM) of the image of 16–21 ∘ was achieved.

Journal of InstrumentationVol. 20(02)
University of Geneva (CH), Czech Technical University in Prague (CZ)
Openalex Percentile: Top 9%
Particle Detector Development and Performance
3.74
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.