Rad-hard readout system for Timepix3 Hybrid Pixel Detectors
Abstract The Beam Gas Ionisation (BGI) profile monitor, located in the Proton Synchrotron (PS) and Super Proton Synchrotron (SPS) at CERN, requires a radiation-tolerant readout system to transfer data from the challenging accelerator surroundings to the back-end for processing. The system needs to control and acquire data from four Timepix3 Hybrid Pixel Detectors (HPDs) located directly inside the beam pipe, a highly radioactive environment. It must ensure reliability given limited hardware access and preserve signal integrity for the high-speed data (32 channels at 320 MHz). However, due to the unavailability of a suitable rad-hard Timepix3 readout, the Beam Instrumentation PiXeL (BIPXL) readout system was designed to meet these requirements. This system employs radiation-hardened components such as the GBTx and the FEASTMP, both developed at CERN. It will be compatible with forthcoming hybrid pixel detector initiatives in similarly harsh radiation conditions.
Authors
- G. G. Cabrera
- Hampus Sandberg
- Mark McLean (ORCID: https://orcid.org/0000-0003-0336-5972)
- Steen Jensen
- James Storey
- Juri Joul
Institutions
- European Organization for Nuclear Research (CH)
Publication Details
- Journal
- Journal of Instrumentation
- Published
- 2025-02-01
- DOI
- https://doi.org/10.1088/1748-0221/20/02/c02009
- Citations
- 1
- Primary Topic
- CCD and CMOS Imaging Sensors
- Type
- article
- Field-Weighted Citation Impact
- 0.73