Impact of porous lung substitute on linear energy transfer (LET) assessed via Monte Carlo simulation and CR‐39 measurement with a carbon‐ion beam
Porous lung-equivalent structures significantly affect LET<sub>d</sub> distributions in carbon-ion therapy, as confirmed by both CR-39 measurements and MC simulations. IMPT with LET<sub>d</sub> optimization may be more impacted by porous structures in terms of median and minimum LET<sub>d</sub> values within the target. The Gaussian convolution function shows promise for enhancing LET<sub>d</sub> calculation accuracy, but further validation in anatomically complex models is needed to assess its clinical feasibility.