Contact Detection of Forceps for a Vitreoretinal Surgical Simulator Using Infrared Reflection Images of Forceps

To train for vitreoretinal surgeries, such as peeling of the inner limiting membrane, a surgical simulator with an elaborate retinal model and method of contact detection is necessary. Quantitative evaluation of forceps contacts to the retinal model is required because excessive forceps contact can cause loss of eyesight. This paper presents a detection method using image analysis to detect forceps contact with the retinal model. An evaluation system for contact detection of forceps was constructed, which includes an optical system incorporating an 850 nm LED, a CCD camera, and an infrared pass filter placed under a retinal model. This system observes the reflected light of infrared light at the forceps tips and evaluates the variation of the forceps shape. The luminance value of the forceps tips edge was set as a common threshold used for each frame and the number of pixels with luminance higher than the threshold was counted before and after forceps contact. As a result of image analysis, we confirmed that the number of counted pixels increased immediately after forceps contact and by evaluating the change in the number of counted pixels, we succeeded in detecting forceps contact at a load of less than 0.1 mN within a radius of $2 \mu \mathrm{~m}$.

Authors

Institutions

Publication Details

Published
2024-11-20
DOI
https://doi.org/10.1109/cbs61689.2024.10860579
Primary Topic
Surgical Simulation and Training
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Contact Detection of Forceps for a Vitreoretinal Surgical Simulator Using Infrared Reflection Images of Forceps

Kuniaki Noda, Hisataka Maruyama
Surgical Simulation and Training
article

Contact Detection of Forceps for a Vitreoretinal Surgical Simulator Using Infrared Reflection Images of Forceps

Kuniaki Noda, Hisataka Maruyama
article en

Abstract

To train for vitreoretinal surgeries, such as peeling of the inner limiting membrane, a surgical simulator with an elaborate retinal model and method of contact detection is necessary. Quantitative evaluation of forceps contacts to the retinal model is required because excessive forceps contact can cause loss of eyesight. This paper presents a detection method using image analysis to detect forceps contact with the retinal model. An evaluation system for contact detection of forceps was constructed, which includes an optical system incorporating an 850 nm LED, a CCD camera, and an infrared pass filter placed under a retinal model. This system observes the reflected light of infrared light at the forceps tips and evaluates the variation of the forceps shape. The luminance value of the forceps tips edge was set as a common threshold used for each frame and the number of pixels with luminance higher than the threshold was counted before and after forceps contact. As a result of image analysis, we confirmed that the number of counted pixels increased immediately after forceps contact and by evaluating the change in the number of counted pixels, we succeeded in detecting forceps contact at a load of less than 0.1 mN within a radius of $2 \mu \mathrm{~m}$.

Nagoya University (JP)
Openalex Percentile: Top 54%
Surgical Simulation and Training
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.