A Novel Patent Foramen Ovale Occluder With Left Clamp Arms

Traditional occluders for the closure of patent foramen ovale (PFO) may be associated with some potential risks, including device-related thrombus, induction of arrhythmias, residual shunt, and difficulty in puncturing atrial septum. The TriFlower device is a novel transcatheter PFO occluder that features a clamp arm design. The device success rate was 100% (n = 18) in the animal experiment. Endothelialization of the left clamp arms was completed at an early stage (30 days after implantation). The acute procedure success rate in 17 patients was 100%. At 6-month follow-up, the primary efficacy endpoint, an effective PFO closure defined as residual shunt grade 1 or lower, was met in 16 (94.1%) patients and 14 (82.4%) patients had a complete closure. There were no device-related complications. These findings indicate that the TriFlower device is feasible, safe, and effective for transcatheter PFO closure. Its potential advantages include rapid endothelialization and low left atrial interference.

Authors

Institutions

Publication Details

Journal
JACC Basic to Translational Science
Published
2025-02-05
DOI
https://doi.org/10.1016/j.jacbts.2024.11.012
Citations
1
Primary Topic
Cardiovascular and Diving-Related Complications
Type
article
Field-Weighted Citation Impact
1.52

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Novel Patent Foramen Ovale Occluder With Left Clamp Arms

Shiqiang Hou, Daxin Zhou, Yuliang Long, Yuan Zhang et al.
1 citations
JACC Basic to Translational Science
Cardiovascular and Diving-Related Complications
1.52
article

A Novel Patent Foramen Ovale Occluder With Left Clamp Arms

Shiqiang Hou, Daxin Zhou, Yuliang Long, Yuan Zhang, Peng Wang, Xiaochun Zhang, Lei Zhang, Mingfei Li, Junbo Ge, Wenzhi Pan
article en
1 citations

Abstract

Traditional occluders for the closure of patent foramen ovale (PFO) may be associated with some potential risks, including device-related thrombus, induction of arrhythmias, residual shunt, and difficulty in puncturing atrial septum. The TriFlower device is a novel transcatheter PFO occluder that features a clamp arm design. The device success rate was 100% (n = 18) in the animal experiment. Endothelialization of the left clamp arms was completed at an early stage (30 days after implantation). The acute procedure success rate in 17 patients was 100%. At 6-month follow-up, the primary efficacy endpoint, an effective PFO closure defined as residual shunt grade 1 or lower, was met in 16 (94.1%) patients and 14 (82.4%) patients had a complete closure. There were no device-related complications. These findings indicate that the TriFlower device is feasible, safe, and effective for transcatheter PFO closure. Its potential advantages include rapid endothelialization and low left atrial interference.

JACC Basic to Translational ScienceVol. 10(4)
Fudan University (CN)
Shanghai Science and Technology Development Foundation, National Natural Science Foundation of China, Ministry of Science and Technology of the People's Republic of China, Fudan University, Science and Technology Commission of Shanghai Municipality
Openalex Percentile: Top 22%
Cardiovascular and Diving-Related Complications
1.52
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A Novel Patent Foramen Ovale Occluder With Left Clamp Arms — Shiqiang Hou, Daxin Zhou, et al. · JACC Basic to Translational Science (2025) | TGRS Research Map | TGRS