A model-independent method for mitigation of pileup artefacts in energy-resolved radiation counting

This article presents a model-independent post-processing method to restore the counting and spectral information of pulse-height analyzers in the presence of pulse pileup at high counting rates. The method is based on the measurement of counts and of the integral durations of the detector signals over different thresholds, and on the comparison of the measurements from two sources with the same energy distribution and different counting rates ρ . The procedure does not require the modeling of the pulse shape, the knowledge of the dead-time τ or other parameters. The correction technique was validated with simulations and with spectra of high activity gamma sources collected with a high-purity germanium (HPGe) detector. For a simulated flat energy distribution and a ratio 2 between the two counting rates, the relative mean deviation between the corrected and input spectra is less than 1% and the relative standard deviation below 5% up to a normalized input rate ρ τ = 1 . 0 . These results improve by increasing the difference between the input rates of the two sources. In the simulation of an X-ray spectrum at a normalized input rate ρ τ = 0 . 5 , the ratio between the number of corrected and generated counts in each energy bin is below a few percent. The experimental validation was performed by acquiring two spectra of 241 Am sources of different activities placed in a fixed position and of a high activity 137 Cs source placed in two different position. In both cases, the correction method provides count rate distributions in close agreement with a reference spectrum from a lower activity source. The mitigation of artefacts in HPGe spectra overcomes the performance of a standard pileup rejection method. • Mitigation of pileup spectrum distortions in energy-resolved radiation counting. • Based on two-sources with the same energy distribution and different counting rates. • Results independent of models or parameterizations. • Better performance than standard pileup rejection methods.

Authors

Institutions

Publication Details

Journal
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Published
2025-02-05
DOI
https://doi.org/10.1016/j.nima.2025.170276
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A model-independent method for mitigation of pileup artefacts in energy-resolved radiation counting

V. Monti, L. Visca, V. Monaco, E. Durisi
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Radiation Detection and Scintillator Technologies
article

A model-independent method for mitigation of pileup artefacts in energy-resolved radiation counting

V. Monti, L. Visca, V. Monaco, E. Durisi
article en

Abstract

This article presents a model-independent post-processing method to restore the counting and spectral information of pulse-height analyzers in the presence of pulse pileup at high counting rates. The method is based on the measurement of counts and of the integral durations of the detector signals over different thresholds, and on the comparison of the measurements from two sources with the same energy distribution and different counting rates ρ . The procedure does not require the modeling of the pulse shape, the knowledge of the dead-time τ or other parameters. The correction technique was validated with simulations and with spectra of high activity gamma sources collected with a high-purity germanium (HPGe) detector. For a simulated flat energy distribution and a ratio 2 between the two counting rates, the relative mean deviation between the corrected and input spectra is less than 1% and the relative standard deviation below 5% up to a normalized input rate ρ τ = 1 . 0 . These results improve by increasing the difference between the input rates of the two sources. In the simulation of an X-ray spectrum at a normalized input rate ρ τ = 0 . 5 , the ratio between the number of corrected and generated counts in each energy bin is below a few percent. The experimental validation was performed by acquiring two spectra of 241 Am sources of different activities placed in a fixed position and of a high activity 137 Cs source placed in two different position. In both cases, the correction method provides count rate distributions in close agreement with a reference spectrum from a lower activity source. The mitigation of artefacts in HPGe spectra overcomes the performance of a standard pileup rejection method. • Mitigation of pileup spectrum distortions in energy-resolved radiation counting. • Based on two-sources with the same energy distribution and different counting rates. • Results independent of models or parameterizations. • Better performance than standard pileup rejection methods.

Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated EquipmentVol. 1073
Istituto Nazionale di Fisica Nucleare, Sezione di Torino (IT), University of Turin (IT)
Climate action
Openalex Percentile: Top 97%
Radiation Detection and Scintillator Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.