Quantifying radiotherapy beam quality: an analysis using gamma passing rates

<i>Purpose</i>. PDD and profile curves play a crucial role in analyzing the beam quality and energy stability of accelerators. The aim of this study was to assess the efficacy of GPR in machine QA and compare it with traditional methods for analyzing dose outputs.<i>Methods</i>. GPRs were employed to assess the quality of radiation beams by comparing 1D and 2D Profile metrics and PDD data against commissioning data. The data used were obtained from the ASCII data files derived from the water tank. GPRs were calculated for all plots with a lower percentage dose cutoff of 10%. The local GPRs and dose influence for the 2D PDD metrics and dose influence were calculated for an open field 10 × 10 cm<sup>2</sup>photon beam at SSD = 100 cm. In both 1D and 2D GPRs analyses, criterion of 1%/1 mm was adopted, as this approach allows for the capture of more subtle variations in the data. To substantiate the viability of the study, a comparative analysis was conducted by comparing the outcomes of the gamma analysis with those derived from traditional methods, such as manual machine quality assurance checks.<i>Results</i>. GPRs demonstrated a superior capability for comprehensive data analysis compared to traditional methods. For the 1D curves, the passing rates (<i>γ</i>≤ 1) are 96.19%, 100%, and 93.46%, respectively. With respect to the 2D dose influence, the PDD image passing rate was 99.57%, and significant dose differences were observed at the four corners of the open field, indicating areas that require further investigation.<i>Conclusions</i>. Compared to traditional methods, GPRs are more sensitive to subtle changes in the data, providing valuable insights into the accelerator beam status.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2025-02-05
DOI
https://doi.org/10.1088/2057-1976/adb291
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quantifying radiotherapy beam quality: an analysis using gamma passing rates

Lirong Fu, Sijia Chen, Yingrong Yu -, Yipeng He et al.
Biomedical Physics & Engineering Express
Advanced Radiotherapy Techniques
article

Quantifying radiotherapy beam quality: an analysis using gamma passing rates

Lirong Fu, Sijia Chen, Yingrong Yu -, Yipeng He, Liwan Shi, Kang Zheng, Xiang Gao, Guanglu Gao
article en

Abstract

<i>Purpose</i>. PDD and profile curves play a crucial role in analyzing the beam quality and energy stability of accelerators. The aim of this study was to assess the efficacy of GPR in machine QA and compare it with traditional methods for analyzing dose outputs.<i>Methods</i>. GPRs were employed to assess the quality of radiation beams by comparing 1D and 2D Profile metrics and PDD data against commissioning data. The data used were obtained from the ASCII data files derived from the water tank. GPRs were calculated for all plots with a lower percentage dose cutoff of 10%. The local GPRs and dose influence for the 2D PDD metrics and dose influence were calculated for an open field 10 × 10 cm<sup>2</sup>photon beam at SSD = 100 cm. In both 1D and 2D GPRs analyses, criterion of 1%/1 mm was adopted, as this approach allows for the capture of more subtle variations in the data. To substantiate the viability of the study, a comparative analysis was conducted by comparing the outcomes of the gamma analysis with those derived from traditional methods, such as manual machine quality assurance checks.<i>Results</i>. GPRs demonstrated a superior capability for comprehensive data analysis compared to traditional methods. For the 1D curves, the passing rates (<i>γ</i>≤ 1) are 96.19%, 100%, and 93.46%, respectively. With respect to the 2D dose influence, the PDD image passing rate was 99.57%, and significant dose differences were observed at the four corners of the open field, indicating areas that require further investigation.<i>Conclusions</i>. Compared to traditional methods, GPRs are more sensitive to subtle changes in the data, providing valuable insights into the accelerator beam status.

Biomedical Physics & Engineering ExpressVol. 11(2)
First Affiliated Hospital of Xiamen University (CN), Zhangzhou Municipal Hospital of Fujian Province (CN), Zhangzhou Vocational and Technical College (CN)
Affordable and clean energy
Openalex Percentile: Top 97%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.