Energy resolution values of GAGG:Ce crystals coupled to various SiPMs

Abstract GAGG:Ce (Gadolinium Aluminum Gallium Garnet doped with Cerium) crystals coupled with SiPMs (Silicon Photomultipliers) form a scintillation detector system commonly used in various fields such as medical imaging, high-energy physics, and radiation detection. The purpose of this study is to compare the energy resolution performance of GAGG:Ce crystals coupled to various SiPMs. Comparisons were conducted among crystals and photomultipliers from various companies, all featuring a pixel size of 3 × 3 mm 2 . The detector system consisted of a crystal optically coupled to a SiPM array, along with the photomultiplier electronics. The entire system was housed within a black box, positioned adjacent to a radiation source—a closed-type 137 Cs isotope, intended for laboratory use—with a radioactivity value of 0.03 MBq. All crystals underwent surface polishing and were subsequently wrapped with multiple layers of Teflon. Additionally, an optical grease was applied between the crystal and the photomultiplier to enhance their optical properties. The size of the crystals and their coupling with the SiPMs were found to significantly impact the energy resolution. Excellent energy resolution values, up to 4.9%, were achieved for the majority of the crystals and SiPMs. Utilizing the 30035 C-Series SiPM from Onsemi, GAGG-HR single crystals exhibited the most favorable energy resolution values, achieving 5% for both 3 × 3 × 5 mm 3 and 3 × 3 × 8 mm 3 sizes. Furukawa GAGG:Ce crystals demonstrated consistent energy resolution ranging from 6.6% to 7.2%. Employing the PM3350 Trench type SiPM by Ketek, GAGG-HR single crystals achieved the best energy resolution values of 5.5% for 3 × 3 × 5 mm 3 and 5.7% for 3 × 3 × 8 mm 3 sizes. Furukawa GAGG:Ce samples of 3 × 3 × 5 mm 3 crystals yielded an energy resolution value of 6.8%. When using the S10362-33 Series SiPM from Hamamatsu, the best energy resolution values were obtained with the S10362-33-050 SiPM coupled with 3 × 3 × 8 mm 3 GAGG:HR crystals, resulting in an energy resolution of 8%.

Authors

Institutions

Publication Details

Journal
The European Physical Journal Special Topics
Published
2025-02-05
DOI
https://doi.org/10.1140/epjs/s11734-025-01477-3
Citations
1
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
1.88

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Energy resolution values of GAGG:Ce crystals coupled to various SiPMs

S. David, Potiriadis Nikolaos, Skouroliakou Aikaterini, Liaparinos Panagiotis
1 citations
The European Physical Journal Special Topics
Radiation Detection and Scintillator Technologies
1.88
article

Energy resolution values of GAGG:Ce crystals coupled to various SiPMs

S. David, Potiriadis Nikolaos, Skouroliakou Aikaterini, Liaparinos Panagiotis
article en
1 citations

Abstract

Abstract GAGG:Ce (Gadolinium Aluminum Gallium Garnet doped with Cerium) crystals coupled with SiPMs (Silicon Photomultipliers) form a scintillation detector system commonly used in various fields such as medical imaging, high-energy physics, and radiation detection. The purpose of this study is to compare the energy resolution performance of GAGG:Ce crystals coupled to various SiPMs. Comparisons were conducted among crystals and photomultipliers from various companies, all featuring a pixel size of 3 × 3 mm 2 . The detector system consisted of a crystal optically coupled to a SiPM array, along with the photomultiplier electronics. The entire system was housed within a black box, positioned adjacent to a radiation source—a closed-type 137 Cs isotope, intended for laboratory use—with a radioactivity value of 0.03 MBq. All crystals underwent surface polishing and were subsequently wrapped with multiple layers of Teflon. Additionally, an optical grease was applied between the crystal and the photomultiplier to enhance their optical properties. The size of the crystals and their coupling with the SiPMs were found to significantly impact the energy resolution. Excellent energy resolution values, up to 4.9%, were achieved for the majority of the crystals and SiPMs. Utilizing the 30035 C-Series SiPM from Onsemi, GAGG-HR single crystals exhibited the most favorable energy resolution values, achieving 5% for both 3 × 3 × 5 mm 3 and 3 × 3 × 8 mm 3 sizes. Furukawa GAGG:Ce crystals demonstrated consistent energy resolution ranging from 6.6% to 7.2%. Employing the PM3350 Trench type SiPM by Ketek, GAGG-HR single crystals achieved the best energy resolution values of 5.5% for 3 × 3 × 5 mm 3 and 5.7% for 3 × 3 × 8 mm 3 sizes. Furukawa GAGG:Ce samples of 3 × 3 × 5 mm 3 crystals yielded an energy resolution value of 6.8%. When using the S10362-33 Series SiPM from Hamamatsu, the best energy resolution values were obtained with the S10362-33-050 SiPM coupled with 3 × 3 × 8 mm 3 GAGG:HR crystals, resulting in an energy resolution of 8%.

The European Physical Journal Special Topics
University of West Attica (GR)
Hellenic Academic Libraries Link, University of West Attica
Openalex Percentile: Top 21%
Radiation Detection and Scintillator Technologies
1.88
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.