Penetration of the Scalp by the Bent Fragment of a Pre-installed Titanium Mini-Plate Due to a Minor Head Injury: A Case Report

Titanium mini-plates are commonly used for bone flap fixation in craniotomy and are particularly essential for covering burr holes. Plate exposure through the scalp may occur because of scalp thinning caused by infection or local ischemia, and penetration of the scalp by a titanium mini-plate that had been bent by a minor head injury are rare. A 69-year-old man who had undergone covering of a burr hole in the calvarium by a titanium plate for clipping of a ruptured aneurysm 17 years ago was referred to our clinic to examine metal protruding through the scalp. One year previously, he had hit his head on a wall with no consequence until one week ago, when he noted the protruding metal. Plain skull radiography showed that a portion of the pinwheel-shaped titanium mini-plate had been bent upward and was exposed through the scalp. As the wound was contaminated with sebum, the plate and three screws securing it were removed through a linear skin incision under local anesthesia. The wound healed uneventfully, and the sutures were removed 10 days later. Bone flap fixation requires careful attention. When a five-screw hole pinwheel-shaped titanium mini-plate is used to cover a burr hole, all five holes should be fixed to the skull/bone flap to prevent bending in the future. Alternatively, a wing that is not fixed with a screw should be cut or a rectangular or diamond-shaped plate should be used.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2025-02-04
DOI
https://doi.org/10.7759/cureus.78472
Primary Topic
Facial Trauma and Fracture Management
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Penetration of the Scalp by the Bent Fragment of a Pre-installed Titanium Mini-Plate Due to a Minor Head Injury: A Case Report

Shoko M. Yamada, S. Nakajima, Yoshihiro Hashimoto
1 links
Cureus
Facial Trauma and Fracture Management
article

Penetration of the Scalp by the Bent Fragment of a Pre-installed Titanium Mini-Plate Due to a Minor Head Injury: A Case Report

Shoko M. Yamada, S. Nakajima, Yoshihiro Hashimoto
article en

Abstract

Titanium mini-plates are commonly used for bone flap fixation in craniotomy and are particularly essential for covering burr holes. Plate exposure through the scalp may occur because of scalp thinning caused by infection or local ischemia, and penetration of the scalp by a titanium mini-plate that had been bent by a minor head injury are rare. A 69-year-old man who had undergone covering of a burr hole in the calvarium by a titanium plate for clipping of a ruptured aneurysm 17 years ago was referred to our clinic to examine metal protruding through the scalp. One year previously, he had hit his head on a wall with no consequence until one week ago, when he noted the protruding metal. Plain skull radiography showed that a portion of the pinwheel-shaped titanium mini-plate had been bent upward and was exposed through the scalp. As the wound was contaminated with sebum, the plate and three screws securing it were removed through a linear skin incision under local anesthesia. The wound healed uneventfully, and the sutures were removed 10 days later. Bone flap fixation requires careful attention. When a five-screw hole pinwheel-shaped titanium mini-plate is used to cover a burr hole, all five holes should be fixed to the skull/bone flap to prevent bending in the future. Alternatively, a wing that is not fixed with a screw should be cut or a rectangular or diamond-shaped plate should be used.

CureusVol. 17(2)
Shizuoka University of Welfare (JP), Teikyo University (JP)
Openalex Percentile: Top 96%
Facial Trauma and Fracture Management
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.