Performance Evaluation of the Ge Discovery IQ PET/CT System According to the Nema NU2-2012 Standard

A PET/CT Scanner -Discovery IQ 5 ring. GE Medical system was installed at Evercare Hospital Dhaka. Total of 11252 BGO crystal elements are arranged in 5 circular rings providing 260 mm axial and 700 mm trans axial field of view. The aim of the test was to assess the performances of the PET/CT scanner GE-Discovery IQ 5 ring. Spatial resolution, Image Quality, and Sensitivity tests were carried out as per NEMA NU2-2012 Standard guidelines. Spatial resolution for radial, tangential and axial directions was carried out at (x,y): (0,1), (0,10), and (0,20) cm locations. Sensitivity was calculated from the activity concentration factor. Image reconstruction algorithm VPHD was used. The radial, tangential, and axial FWHM 4.2 mm, 4.74 mm, and 4.82 mm at 1 cm; 5.72 mm, 4.82 mm, and 4.57 mm at 10 cm off center; 8.17 mm, 4.85 mm, and 4.92 mm at 10 cm off center respectively. In NEMA image quality, the hot contrast recovery values for 10-13-17-22 mm spheres were 25.7- 50.3-56.1-69.7 % respectively and cold contrast recovery values for 28 mm and 37 mm spheres were 75.5 and 77.7 % respectively; the corresponding background variability values were 7.2-6.9-6.6-5.9-5.5-4.7 %. The lung error residual mean was 16 %. Overall, PET performances of the Discovery IQ whole body scanner were satisfactory and passed the entire NEMA NU2- 2012 acceptance test. Bangladesh Journal of Physics, Vol. 30, Issue 2, pp. 57–67, December 2023

Authors

Institutions

Publication Details

Journal
Bangladesh Journal of Physics
Published
2025-02-04
DOI
https://doi.org/10.3329/bjphy.v30i2.68384
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance Evaluation of the Ge Discovery IQ PET/CT System According to the Nema NU2-2012 Standard

Tanvir Ahmed Biman, Md Shohag Mia, Mohammed Fazlul Kabir, Md Nahid Hossain
Bangladesh Journal of Physics
Medical Imaging Techniques and Applications
article

Performance Evaluation of the Ge Discovery IQ PET/CT System According to the Nema NU2-2012 Standard

Tanvir Ahmed Biman, Md Shohag Mia, Mohammed Fazlul Kabir, Md Nahid Hossain
article en

Abstract

A PET/CT Scanner -Discovery IQ 5 ring. GE Medical system was installed at Evercare Hospital Dhaka. Total of 11252 BGO crystal elements are arranged in 5 circular rings providing 260 mm axial and 700 mm trans axial field of view. The aim of the test was to assess the performances of the PET/CT scanner GE-Discovery IQ 5 ring. Spatial resolution, Image Quality, and Sensitivity tests were carried out as per NEMA NU2-2012 Standard guidelines. Spatial resolution for radial, tangential and axial directions was carried out at (x,y): (0,1), (0,10), and (0,20) cm locations. Sensitivity was calculated from the activity concentration factor. Image reconstruction algorithm VPHD was used. The radial, tangential, and axial FWHM 4.2 mm, 4.74 mm, and 4.82 mm at 1 cm; 5.72 mm, 4.82 mm, and 4.57 mm at 10 cm off center; 8.17 mm, 4.85 mm, and 4.92 mm at 10 cm off center respectively. In NEMA image quality, the hot contrast recovery values for 10-13-17-22 mm spheres were 25.7- 50.3-56.1-69.7 % respectively and cold contrast recovery values for 28 mm and 37 mm spheres were 75.5 and 77.7 % respectively; the corresponding background variability values were 7.2-6.9-6.6-5.9-5.5-4.7 %. The lung error residual mean was 16 %. Overall, PET performances of the Discovery IQ whole body scanner were satisfactory and passed the entire NEMA NU2- 2012 acceptance test. Bangladesh Journal of Physics, Vol. 30, Issue 2, pp. 57–67, December 2023

Bangladesh Journal of PhysicsVol. 30(2)
Dhaka Community Hospital Trust (BD), National Institute of Nuclear Medicine & Allied Sciences (BD)
Openalex Percentile: Top 97%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Performance Evaluation of the Ge Discovery IQ PET/CT System According to the Nema NU2-2012 Standard — Tanvir Ahmed Biman, Md Shohag Mia, et al. · Bangladesh Journal of Physics (2025) | TGRS Research Map | TGRS