Monte Carlo toolkit for designing and validating step-range-filter spectrometer designs

Here, we present a Monte Carlo toolkit for validating step range filter (SRF) spectrometer designs. Geant4 is used to transport charged particles through the SRF filters to generate synthetic SRF data that include realistic CR-39 effects. Synthetic SRF spectra generated by this method inherently account for instrument response and allow for the quantification of SRF performance before shots. The usefulness of this toolkit is demonstrated through its application to a number of problems. A new broadband SRF for the ∼10 MeV wide 3He3He proton spectrum is validated, and an analysis method for analyzing 3He3He-p SRF data that accounts for instrument response is put forth. In addition, an SRF design for the compact recoil-proton spectrometer (CRS) on the Z-machine is validated. Finally, a new calibration technique for the DD-p SRF is proposed and validated.

Authors

Institutions

Publication Details

Journal
Review of Scientific Instruments
Published
2025-02-01
DOI
https://doi.org/10.1063/5.0206206
Citations
1
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
1.88

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Monte Carlo toolkit for designing and validating step-range-filter spectrometer designs

P. J. Adrian, Maria Gatu Johnson, S. G. Dannhoff, J. A. Pearcy et al.
1 citations
Review of Scientific Instruments
Radiation Detection and Scintillator Technologies
1.88
article

Monte Carlo toolkit for designing and validating step-range-filter spectrometer designs

P. J. Adrian, Maria Gatu Johnson, S. G. Dannhoff, J. A. Pearcy, C. Chang, Chikang Li, T. E. Evans, R. D. Petrasso, B. Lahmann, B. L. Reichelt, A. Birkel, Johan A. Frenje, T. M. Johnson, F. H. Séguin, M. Cufari, J. H. Kunimune, N. V. Kabadi, G. D. Sutcliffe, L. Russell, N. Vanderloo
article en
1 citations

Abstract

Here, we present a Monte Carlo toolkit for validating step range filter (SRF) spectrometer designs. Geant4 is used to transport charged particles through the SRF filters to generate synthetic SRF data that include realistic CR-39 effects. Synthetic SRF spectra generated by this method inherently account for instrument response and allow for the quantification of SRF performance before shots. The usefulness of this toolkit is demonstrated through its application to a number of problems. A new broadband SRF for the ∼10 MeV wide 3He3He proton spectrum is validated, and an analysis method for analyzing 3He3He-p SRF data that accounts for instrument response is put forth. In addition, an SRF design for the compact recoil-proton spectrometer (CRS) on the Z-machine is validated. Finally, a new calibration technique for the DD-p SRF is proposed and validated.

Review of Scientific InstrumentsVol. 96(2)
Lawrence Livermore National Laboratory (US), Los Alamos National Laboratory (US), Energetics (United States) (US), University of Rochester (US), Massachusetts Institute of Technology (US)
National Nuclear Security Administration
Openalex Percentile: Top 21%
Radiation Detection and Scintillator Technologies
1.88
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.