Commutability study on three CRMs evaluating their suitability for calibration, trueness verification and statistical quality control of methods measuring metal concentrations in human blood

The commutability of three certified reference materials for Cd, Cr, Hg, Ni, Pb and Tl in blood was evaluated for the combination of digestion ICP-MS with dilution ICP-MS or dilution GFAAS using the difference in bias approach. Maximum non-commutability bias (MANCB) was derived from the median of the tolerance limits set by six proficiency testing (PT) providers. The evaluation showed that this criteria are too strict and that either unfeasible high number of measurements or a larger MANCB is required to demonstrate commutability. The effect of increasing the MANCB on the maximum tolerable measurement uncertainty was found small to moderate demonstrating the commutability of the materials when used as a calibrator. Also the inclusion of the uncertainty of non-commutability into the measurement uncertainty resulted in only small increases of the measurement uncertainty, demonstrating the suitability of the materials for trueness control and method validation. The bias of the three CRMs was for most elements within the 99% confidence limits of the bias of the patient samples, demonstrating the suitability of the materials for statistical quality control.

Authors

Institutions

Publication Details

Journal
Analytical and Bioanalytical Chemistry
Published
2025-02-04
DOI
https://doi.org/10.1007/s00216-025-05751-0
Citations
3
Primary Topic
Pesticide Residue Analysis and Safety
Type
article
Field-Weighted Citation Impact
3.49

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Commutability study on three CRMs evaluating their suitability for calibration, trueness verification and statistical quality control of methods measuring metal concentrations in human blood

Liesbet Deprez, Guy Auclair, Thomas P. J. Linsinger
3 citations
Analytical and Bioanalytical Chemistry
Pesticide Residue Analysis and Safety
3.49
article

Commutability study on three CRMs evaluating their suitability for calibration, trueness verification and statistical quality control of methods measuring metal concentrations in human blood

Liesbet Deprez, Guy Auclair, Thomas P. J. Linsinger
article en
3 citations

Abstract

The commutability of three certified reference materials for Cd, Cr, Hg, Ni, Pb and Tl in blood was evaluated for the combination of digestion ICP-MS with dilution ICP-MS or dilution GFAAS using the difference in bias approach. Maximum non-commutability bias (MANCB) was derived from the median of the tolerance limits set by six proficiency testing (PT) providers. The evaluation showed that this criteria are too strict and that either unfeasible high number of measurements or a larger MANCB is required to demonstrate commutability. The effect of increasing the MANCB on the maximum tolerable measurement uncertainty was found small to moderate demonstrating the commutability of the materials when used as a calibrator. Also the inclusion of the uncertainty of non-commutability into the measurement uncertainty resulted in only small increases of the measurement uncertainty, demonstrating the suitability of the materials for trueness control and method validation. The bias of the three CRMs was for most elements within the 99% confidence limits of the bias of the patient samples, demonstrating the suitability of the materials for statistical quality control.

Analytical and Bioanalytical ChemistryVol. 417(12)
European Commission (BE)
HORIZON EUROPE Non-nuclear direct actions of the Joint Research Centre
Openalex Percentile: Top 10%
Pesticide Residue Analysis and Safety
3.49
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Commutability study on three CRMs evaluating their suitability for calibration, trueness verification and statistical quality control of methods measuring metal concentrations in human blood — Liesbet Deprez, Guy Auclair, et al. · Analytical and Bioanalytical Chemistry (2025) | TGRS Research Map | TGRS